RayStation with the following product descriptions: 1. RayStation 11A, Software Version: 11.0.0.951. 2. RayStation 11A Service Pack 1, Software Version: 11.0.1.29. 3. RayStation 11A Service Pack 2, Software Version: 11.0.3.116. 4. RayStation 11A Service Pack 3, Software Version: 11.0.4.15. Product Description: Radiation Therapy Treatment Planning System.
Published September 10, 2025 · Recall initiated: August 8, 2025
Summary
FDA records medical device Class II recall Z-2491-2025, published September 10, 2025 and initiated August 8, 2025. Recalling firm: RAYSEARCH LABORATORIES AB; Status: Ongoing; Reported quantity: 3; Lot: 1. Software Version: 11.0.0.951, UDI: 0735000201038920210518. 2. Software Version: 11.0.1.29, UDI: 0735000201043320210610. 3. Software Version: 11.0.3.116, UDI: 0735000201044020210916. 4. Software Version: 11.0.4.15, UDI: 0735000201063120220616. Expiration Date: 2027-06-22. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Distributor
- Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL, IA, KY, LA, MD, MN, MS, NY, NC, OH, OK, SC, WA, WI and the countries of Canada, China, Colombia, France, Germany, Hong Kong S.A.R., China, Italy, Japan, Netherlands, Poland, Spain, Sweden, Switzerland, ENG, Thailand, Turkey, UK, Ukraine.
- Lot
- 1. Software Version: 11.0.0.951, UDI: 0735000201038920210518. 2. Software Version: 11.0.1.29, UDI: 0735000201043320210610. 3. Software Version: 11.0.3.116, UDI: 0735000201044020210916. 4. Software Version: 11.0.4.15, UDI: 0735000201063120220616. Expiration Date: 2027-06-22
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Sweden
Related recalls
- FDARayCare 2024A SP4. Oncology Information Systems.July 8, 2026
- FDARayCare 2024A SP1. Oncology Information Systems.July 8, 2026
- FDABrand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 14.0.0, 15.0.0, 15.1.3 SoftwareJanuary 28, 2026
- FDABrand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version:January 28, 2026
- FDARayStation with the following product descriptions: 1. RayStation 9B, Software Version: 9.1.0.933. 2.September 10, 2025
- FDAInPen App, CFN: MMT-8060 (iOS Users) and MMT-8061 (Android Users)September 10, 2025
Questions about this recall
Was RayStation with the following product descriptions: 1. RayStation 11A, Software Version: 11.0.0.951. 2. RayStation 11A Service Pack 1, Software Version: 11.0.1.29. 3. RayStation 11A Service Pack 2, Software Version: 11.0.3.116. 4. RayStation 11A Service Pack 3, Software Version: 11.0.4.15. Product Description: Radiation Therapy Treatment Planning System. recalled?
Yes. FDA published recall Z-2491-2025 on September 10, 2025. Reported hazard: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
What should I do if I have this product?
Voluntary: Firm initiated · Letter