Philips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A
Published July 1, 2026 · Recall initiated: May 26, 2026
Summary
FDA records medical device Class II recall Z-2522-2026, published July 1, 2026 and initiated May 26, 2026. Recalling firm: Philips North America Llc; Status: Ongoing; Reported quantity: 285 units; Lot: UDI/DI 00884838000407, Serial Numbers: DE79569306, DE79569332, DE79569360, DE79569417, DE79569442, DE79569482, DE79569307, DE79569470, DE79569390, DE79569356, DE79569452, DE79569289, DE79569302, DE79569536, DE79569358, DE79569343, DE79569457, DE79569353, DE79569373, DE79569411, DE79569491, DE79569615, DE79569287, DE79569338, DE79569437, DE79569459, DE79569473, DE79569340, DE79569308, DE79569377, DE79569286, DE79569304, DE79569327, DE79569330, DE79569347, DE79569352, DE79569400, DE79569418, DE79569431, DE79569436, DE79569456, DE79569469, DE79569474, DE79569588, DE79569326, DE79569424, DE79569487, DE79569547, DE79569290, DE79569258, DE79569367, DE79569370, DE79569380, DE79569388, DE79569311, DE79569440, DE79569460, DE79569527, DE79569462, DE79569299, DE79569261, DE79569405, DE79569260, DE79569313, DE79569369, DE79569216,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips North America Llc
- Distributor
- Worldwide distribution - US Nationwide and The countries of AE, AT, AU, BE, BG, BH, BR, CA, CH, CN, CR, CZ, DE, ES, FI, FR, GB, GR, HK, ID, IL, IN, IT, JO, JP, KR, KR, LT, LU, LV, MX, MY, NL, NZ, OM, SA, SE, SG, SI, TR, TW, ZA.
- Lot
- UDI/DI 00884838000407, Serial Numbers: DE79569306, DE79569332, DE79569360, DE79569417, DE79569442, DE79569482, DE79569307, DE79569470, DE79569390, DE79569356, DE79569452, DE79569289, DE79569302, DE79569536, DE79569358, DE79569343, DE79569457, DE79569353, DE79569373, DE79569411, DE79569491, DE79569615, DE79569287, DE79569338, DE79569437, DE79569459, DE79569473, DE79569340, DE79569308, DE79569377, DE79569286, DE79569304, DE79569327, DE79569330, DE79569347, DE79569352, DE79569400, DE79569418, DE79569431, DE79569436, DE79569456, DE79569469, DE79569474, DE79569588, DE79569326, DE79569424, DE79569487, DE79569547, DE79569290, DE79569258, DE79569367, DE79569370, DE79569380, DE79569388, DE79569311, DE79569440, DE79569460, DE79569527, DE79569462, DE79569299, DE79569261, DE79569405, DE79569260, DE79569313, DE79569369, DE79569216,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333October 7, 2026
- FDAAchieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (AchievaSeptember 30, 2026
- FDAIngenia 3.0T and Ingenia 3.0T CX; Model Numbers (REF): 1. 781342(3.0T), 2. 781377(3.0T), 3. 782103(3.0T), 4.September 30, 2026
- FDAIngenia 1.5T, Ingenia 1.5T CX and Ingenia 1.5T S; Model Numbers (REF): 1. 781315(1.5T), 2. 781341(1.5T), 3.September 30, 2026
- FDADa Vinci ASSY, PSS, SP1098, Part Number: 380601July 1, 2026
- FDAPhilips Avalon Fetal Monitor, FM 30 Part numberM2703AJuly 1, 2026
Questions about this recall
Was Philips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A recalled?
Yes. FDA published recall Z-2522-2026 on July 1, 2026. Reported hazard: Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.
What should I do if I have this product?
Voluntary: Firm initiated · Letter