Back to search
FDAFDA · Food and drugs·Recall number Z-2523-2026

Philips Avalon Fetal Monitor, FM 30 Part numberM2703A

Published July 1, 2026 · Recall initiated: May 26, 2026

Summary

FDA records medical device Class II recall Z-2523-2026, published July 1, 2026 and initiated May 26, 2026. Recalling firm: Philips North America Llc; Status: Ongoing; Reported quantity: 169 units; Lot: UDI/DI 00884838000414, Serial Numbers: DE79646555, DE79646616, DE79646645, DE79646647, DE79646651, DE79646697, DE79646702, DE79646746, DE79646754, DE79646757, DE79646758, DE79646762, DE79646765, DE79646775, DE79646776, DE79646783, DE79646857, DE79646891, DE79646924, DE79646750, DE79646751, DE79646756, DE79646804, DE79646592, DE79646877, DE79646644, DE79646706, DE79646723, DE79646653, DE79646766, DE79646786, DE79646824, DE79646827, DE79646848, DE79646956, DE79646709, DE79646892, DE79646724, DE79646729, DE79646732, DE79646733, DE79646927, DE79646675, DE79646820, DE79646821, DE79646837, DE79646866, DE79646869, DE79646871, DE79646875, DE79646882, DE79646683, DE79646676, DE79646640, DE79646849, DE79646952, DE79646699, DE79646694, DE79646648, DE79646701, DE79646900, DE79647027, DE79647039, DE79646654, DE79646800, DE79646604, DE79646660, DE79646829, DE79646747, DE79646606, DE79646826, DE79646784, DE79646794. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips North America Llc
Distributor
Worldwide distribution - US Nationwide and The countries of AE, AT, AU, BE, BG, BH, BR, CA, CH, CN, CR, CZ, DE, ES, FI, FR, GB, GR, HK, ID, IL, IN, IT, JO, JP, KR, KR, LT, LU, LV, MX, MY, NL, NZ, OM, SA, SE, SG, SI, TR, TW, ZA.
Lot
UDI/DI 00884838000414, Serial Numbers: DE79646555, DE79646616, DE79646645, DE79646647, DE79646651, DE79646697, DE79646702, DE79646746, DE79646754, DE79646757, DE79646758, DE79646762, DE79646765, DE79646775, DE79646776, DE79646783, DE79646857, DE79646891, DE79646924, DE79646750, DE79646751, DE79646756, DE79646804, DE79646592, DE79646877, DE79646644, DE79646706, DE79646723, DE79646653, DE79646766, DE79646786, DE79646824, DE79646827, DE79646848, DE79646956, DE79646709, DE79646892, DE79646724, DE79646729, DE79646732, DE79646733, DE79646927, DE79646675, DE79646820, DE79646821, DE79646837, DE79646866, DE79646869, DE79646871, DE79646875, DE79646882, DE79646683, DE79646676, DE79646640, DE79646849, DE79646952, DE79646699, DE79646694, DE79646648, DE79646701, DE79646900, DE79647027, DE79647039, DE79646654, DE79646800, DE79646604, DE79646660, DE79646829, DE79646747, DE79646606, DE79646826, DE79646784, DE79646794.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Philips Avalon Fetal Monitor, FM 30 Part numberM2703A recalled?

Yes. FDA published recall Z-2523-2026 on July 1, 2026. Reported hazard: Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.

What should I do if I have this product?

Voluntary: Firm initiated · Letter