Hookup Model/Catalog Number: 2-8-540HAN Product Description: Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor and Advantage Plus" Pass-Thru Automated Endoscope Reprocessor ( AER ) and endoscopes, which allow perfusion of liquids dispensed by the AER to contact and high-level disinfect the channels of flexible endoscopes.
Published August 14, 2024 · Recall initiated: June 13, 2024
Summary
FDA records medical device Class II recall Z-2537-2024, published August 14, 2024 and initiated June 13, 2024. Recalling firm: Steris Corporation; Status: Ongoing; Reported quantity: 129; Lot: Lot Code: UDI-DI 00677964086885; Lot Numbers 415308, 415665, 422921, 456865, 458947, 461749, 463591, 465821, 467121, 467705, 468813, 480858, 484468, 513285, 517588, 533998, 544724, 585095, 617851. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Steris Corporation
- Distributor
- Worldwide distribution - United States Nationwide and the countries of Canada, Belgium, Czech Republic, Denmark, Germany, Spain, Italy, Slovenia, Slovakia, Australia, Israel, Jordan, Kuwait, Namibia, Saudi Arabia, South Africa, Cote d Ivoire, Afghanistan, Albania, Belarus, Botswana, Kazakhstan, Malaysia, Qatar, Ukraine, United Arab Emirates, Nicaragua, and Nigeria.
- Lot
- Lot Code: UDI-DI 00677964086885; Lot Numbers 415308, 415665, 422921, 456865, 458947, 461749, 463591, 465821, 467121, 467705, 468813, 480858, 484468, 513285, 517588, 533998, 544724, 585095, 617851
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116October 21, 2026
- FDAVERIFY Biological Steam Test Pack Model/Catalog Number: S3069October 14, 2026
- FDAHookup Model/Catalog Number: 2-8-540 Product Description: Hookups provide the interface between an AdvantageAugust 14, 2024
- FDAHookup Model/Catalog Number: 2-8-540CAS Product Description: Hookups provide the interface between anAugust 14, 2024
Questions about this recall
Was Hookup Model/Catalog Number: 2-8-540HAN Product Description: Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor and Advantage Plus" Pass-Thru Automated Endoscope Reprocessor ( AER ) and endoscopes, which allow perfusion of liquids dispensed by the AER to contact and high-level disinfect the channels of flexible endoscopes. recalled?
Yes. FDA published recall Z-2537-2024 on August 14, 2024. Reported hazard: IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection
What should I do if I have this product?
Voluntary: Firm initiated · Letter