Zimmer PERSONA Distal Valgus Alignment Guide, Nonsterile. orthopedic surgical instrument. The device sets the proper Varus/Valgus alignment of a patient's knee prior to the initial femoral resection cut being made.

Published September 3, 2014 · Recall initiated: July 30, 2014

Summary

FDA records medical device Class II recall Z-2543-2014, published September 3, 2014 and initiated July 30, 2014. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 2,248 units; Lot: Item Number 42509900400, Lot Number: 62313324, 62415340, 62561031, 62668655, 62313326, 62430425, 62580270, 62695918, 62341561, 62442314, 62590506, 62702686, 62358628, 62473246, 62594059, 62715932, 62368700, 62507846, 62611157, 62728374, 62376588, 62522771, 62624943, 62742362, 62396348, 62530133, 62637095, 62749057, 62401156, 62534170, 62641120, 62757397, 62407514, 62538842, 62662172, 62766074, 62410673, 62542958, & 62668654. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Persona Distal Valgus Alignment guide collet locks are seizing in the 'unlocked' position.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide Distribution -- United States, including Puerto Rico; and, to the countries of Australia, Austria, Japan, Belgium, Italy, Switzerland, India, Germany, Spain, France, Israel, Netherlands, Saudi Arabia, South Africa & Canada.
Lot
Item Number 42509900400, Lot Number: 62313324, 62415340, 62561031, 62668655, 62313326, 62430425, 62580270, 62695918, 62341561, 62442314, 62590506, 62702686, 62358628, 62473246, 62594059, 62715932, 62368700, 62507846, 62611157, 62728374, 62376588, 62522771, 62624943, 62742362, 62396348, 62530133, 62637095, 62749057, 62401156, 62534170, 62641120, 62757397, 62407514, 62538842, 62662172, 62766074, 62410673, 62542958, & 62668654.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Zimmer PERSONA Distal Valgus Alignment Guide, Nonsterile. orthopedic surgical instrument. The device sets the proper Varus/Valgus alignment of a patient's knee prior to the initial femoral resection cut being made. recalled?

Yes. FDA published recall Z-2543-2014 on September 3, 2014. Reported hazard: Persona Distal Valgus Alignment guide collet locks are seizing in the 'unlocked' position.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit