FDAFDA · Food and drugs·Recall number Z-2544-2014

Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K, MMT-723, MMT-723K, MMT-554, and MMT-754. MiniMed 530G: MMT-551, and MMT-751 Paradigm insulin infusion pumps are intended for use to deliver basal and bolus insulin to persons with diabetes.

Published September 3, 2014 · Recall initiated: March 13, 2014

Summary

FDA records medical device Class II recall Z-2544-2014, published September 3, 2014 and initiated March 13, 2014. Recalling firm: Medtronic MiniMed Inc.; Status: Terminated; Reported quantity: 559,374 units total (444,374 units in US); Lot: Paradigm: MMT-523NAB, MMT-723KNB, MMT-523NAH, MMT-723KNH, MMT -523NAL, MMT-723KNL, MMT-523NAP, MMT-723KNP, MMT-523NAS, MMT-723KNS, MMT-723NAB, MMT-723NAH, MMT-723NAL, MMT-723NAP, MMT-723NAS, MMT-523KNB, MMT-523KNH, MMT-523KNL, MMT-523KNP, MMT-523KNS. MiniMed 530G: MMT-551NAM, MMT-551NAH, MMT-551NAL, MMT-551NAP, MMT-551NAS, MMT-751NAB, MMT-751NAH, MMT-751NAL, MMT-751NAP, MMT-751NAS. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion Pumps because they have received reports that users have accidentally programmed the pump to deliver the maximum bolus amount.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medtronic MiniMed Inc.
Distributor
Worldwide Distribution-Nationwide in US
Lot
Paradigm: MMT-523NAB, MMT-723KNB, MMT-523NAH, MMT-723KNH, MMT -523NAL, MMT-723KNL, MMT-523NAP, MMT-723KNP, MMT-523NAS, MMT-723KNS, MMT-723NAB, MMT-723NAH, MMT-723NAL, MMT-723NAP, MMT-723NAS, MMT-523KNB, MMT-523KNH, MMT-523KNL, MMT-523KNP, MMT-523KNS. MiniMed 530G: MMT-551NAM, MMT-551NAH, MMT-551NAL, MMT-551NAP, MMT-551NAS, MMT-751NAB, MMT-751NAH, MMT-751NAL, MMT-751NAP, MMT-751NAS.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K, MMT-723, MMT-723K, MMT-554, and MMT-754. MiniMed 530G: MMT-551, and MMT-751 Paradigm insulin infusion pumps are intended for use to deliver basal and bolus insulin to persons with diabetes. recalled?

Yes. FDA published recall Z-2544-2014 on September 3, 2014. Reported hazard: Medtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion Pumps because they have received reports that users have accidentally programmed the pump to deliver the maximum bolus amount.

What should I do if I have this product?

Voluntary: Firm initiated · Letter