Monaco The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon and electron treatment plans and displays, on-screen and in hard-copy, two or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups.
Published September 10, 2014 · Recall initiated: August 18, 2014
Summary
FDA records medical device Class II recall Z-2581-2014, published September 10, 2014 and initiated August 18, 2014. Recalling firm: Elekta, Inc.; Status: Terminated; Reported quantity: 928 units; Lot: Versions 3.20.00 and higher. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Incorrect Delivery of Composite VMAT Fields - When DICOM is exporting an arc plan (VMAT, Dynamic Conformal or 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported plan will not match the approved treatment plan.
What to do
Voluntary: Firm initiated · Other
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Elekta, Inc.
- Distributor
- Worldwide distribution: US (nationwide) including states of: AL, AK, AZ, CA, CO, CT, FL, IL, IN, MD, MA, MI, MN, MO, MT, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SD, TN, TX, VA, WA, WI, Puerto Rico; and countries of: Algeria, Angola, Antigua & Barbuda, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Ireland, Israel, Italy, Japan, Kazakhstan, Libya, Lithuania, Malaysia, Malta, Mexico, Morocco, Myanmar, Namibia, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sir Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Ukraine.
- Lot
- Versions 3.20.00 and higher
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: Geneva Product Name: Brachytherapy Applicator Product Description: The device is intended forJuly 15, 2026
- FDABrand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator Product Description: TheJuly 15, 2026
- FDABrand Name: Fletcher CT/MR Shielded Applicator Set Product Name: Intracavitary Remote Afterloading ApplicatorJuly 15, 2026
- FDAMOSAIQ Oncology Information SystemJune 17, 2026
- FDAPERSONA" The Personalized Knee System UC Tibial Articular Surface Provisional Left Size 3-7 CD Top; Left SizeSeptember 10, 2014
- FDANatural-Knee System Patella Bushings. Orthopedic surgical instrument. Part Number 6290-00-690. Per theSeptember 10, 2014
Questions about this recall
Was Monaco The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon and electron treatment plans and displays, on-screen and in hard-copy, two or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups. recalled?
Yes. FDA published recall Z-2581-2014 on September 10, 2014. Reported hazard: Incorrect Delivery of Composite VMAT Fields - When DICOM is exporting an arc plan (VMAT, Dynamic Conformal or 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported plan will not match the approved treatment plan.
What should I do if I have this product?
Voluntary: Firm initiated · Other