CT Goldseal Optima CT 600
Published August 8, 2018 · Recall initiated: May 24, 2018
Summary
FDA records medical device Class II recall Z-2583-2018, published August 8, 2018 and initiated May 24, 2018. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 42 units total; Lot: Serial Number 51941YC3, System ID Number 406434660. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
- Lot
- Serial Number 51941YC3, System ID Number 406434660
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDAVNS Therapy Programming System, Rx Only, Model 3000, v1.0.2.2August 8, 2018
- FDARevolution EVO 3.6, labeled as a. HVY WAUK; b. LONG HINO; c. LONG WAUK; d. MID BJG; e. MID WAUK.August 8, 2018
Questions about this recall
Was CT Goldseal Optima CT 600 recalled?
Yes. FDA published recall Z-2583-2018 on August 8, 2018. Reported hazard: Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
What should I do if I have this product?
Voluntary: Firm initiated · Letter