FDAFDA · Food and drugs·Recall number Z-2583-2018

CT Goldseal Optima CT 600

Published August 8, 2018 · Recall initiated: May 24, 2018

Summary

FDA records medical device Class II recall Z-2583-2018, published August 8, 2018 and initiated May 24, 2018. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 42 units total; Lot: Serial Number 51941YC3, System ID Number 406434660. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE Healthcare, LLC
Distributor
Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
Lot
Serial Number 51941YC3, System ID Number 406434660
Risk classification
Class II
Area (FDA)
device
Country
United States