***Baxter***Buretrol Solution Set***105" (2.7 m)***150 mL Burette***Drip Chamber Filter Valve***3 Injection Sites Male Luer Lock Adapter with***Retractable Collar***Fluid path is Sterile, Nonpyrogenic***Indications for Use: For the administration of fluids from a container into the patient's vascular system through a vascular access device.***
Published September 17, 2014 · Recall initiated: July 14, 2014
Summary
FDA records medical device Class II recall Z-2613-2014, published September 17, 2014 and initiated July 14, 2014. Recalling firm: Baxter Healthcare Corp; Status: Terminated; Reported quantity: 234196; Lot: 2C7564- Buretrol Solution Set with 150mL INTERLINK Burette, lots: DR13E06019 DR13E07017 DR13E07025 DR13I14023 DR13I16010 DR13I16028 DR13I23016 DR13I23024 DR13J03033 DR13J04015 DR13J04023 DR13J08024 DR13J09014 DR13J09022 DR13J24021 DR13J24039 DR13K07024 DR13K08022 DR13K12024 DR13K13022 DR13K18021 DR13K19029 DR13K25026 DR13K26024 DR13L02023 DR13L03039 DR13L10018 DR13L11016 DR13L18011 DR13L18029 DR13L19035 DR14A09052 DR14A10035 DR14A23012 DR14A23020 DR14A29027 DR14A30017 DR14B06049 DR14B07013 DR14B07047 DR14B11023 DR14B13037 DR14B13045 DR14B14019 DR14B20016 DR14B20032 DR14B21014 DR14B26013 DR14B26039 DR14C05015 DR14C05031. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Baxter Healthcare Corporation is voluntarily issuing a recall for specific lots of Buretrol Solution set product codes due to complaints for separation between the burette chamber and the drip chamber. Affected product codes are: 2C7564, 2C8864 and 2H8864.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Baxter Healthcare Corp
- Distributor
- Nationwide Distribution.
- Lot
- 2C7564- Buretrol Solution Set with 150mL INTERLINK Burette, lots: DR13E06019 DR13E07017 DR13E07025 DR13I14023 DR13I16010 DR13I16028 DR13I23016 DR13I23024 DR13J03033 DR13J04015 DR13J04023 DR13J08024 DR13J09014 DR13J09022 DR13J24021 DR13J24039 DR13K07024 DR13K08022 DR13K12024 DR13K13022 DR13K18021 DR13K19029 DR13K25026 DR13K26024 DR13L02023 DR13L03039 DR13L10018 DR13L11016 DR13L18011 DR13L18029 DR13L19035 DR14A09052 DR14A10035 DR14A23012 DR14A23020 DR14A29027 DR14A30017 DR14B06049 DR14B07013 DR14B07047 DR14B11023 DR14B13037 DR14B13045 DR14B14019 DR14B20016 DR14B20032 DR14B21014 DR14B26013 DR14B26039 DR14C05015 DR14C05031
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACOSEAL Surgical Sealant Kit, 8 mL, Product Code: 934072; For use in vascular reconstructions to achieveJune 8, 2016
- FDACOSEAL Surgical Sealant Kit, 4 mL, Product Code: 934071; For use in vascular reconstructions to achieveJune 8, 2016
- FDAHeparin Sodium and 0.9% Sodium Chloride Injection, 1000 USP Heparin Units, 500 mL in a VIAFLEX Plus Container,June 17, 2015
- FDA0.9% Sodium Chloride Injection USP, 100 mL MINI-BAG Plus Container, Rx only, Baxter Healthcare Corporation,April 15, 2015
- FDAIllumina Worklist Manager (IWM) (software v1.0.15), a component of Illumina MiSeqDx Platform. Product Usage:September 17, 2014
- FDA***Baxter***BURETROL Solution Set***105" (2.7 m) 150 mL Burette***Drip Chamber Filter Valve***3 Luer ActivatedSeptember 17, 2014
Questions about this recall
Was ***Baxter***Buretrol Solution Set***105" (2.7 m)***150 mL Burette***Drip Chamber Filter Valve***3 Injection Sites Male Luer Lock Adapter with***Retractable Collar***Fluid path is Sterile, Nonpyrogenic***Indications for Use: For the administration of fluids from a container into the patient's vascular system through a vascular access device.*** recalled?
Yes. FDA published recall Z-2613-2014 on September 17, 2014. Reported hazard: Baxter Healthcare Corporation is voluntarily issuing a recall for specific lots of Buretrol Solution set product codes due to complaints for separation between the burette chamber and the drip chamber. Affected product codes are: 2C7564, 2C8864 and 2H8864.
What should I do if I have this product?
Voluntary: Firm initiated · Letter