WAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320
Published August 21, 2024 · Recall initiated: July 17, 2024
Summary
FDA records medical device Class II recall Z-2617-2024, published August 21, 2024 and initiated July 17, 2024. Recalling firm: Boston Scientific Neuromodulation Corporation; Status: Ongoing; Reported quantity: 77,674; Lot: REF/UDI-DI(GTIN)/Serial Number Range/Expiration Date Range: M365SC12160/8714729985082/100188 - 758920, M365SC12320/8714729985099/100635 - 759404. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Spinal Cord Stimulation (SCS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset, which may lead to transient loss of stimulation; patients may experience undesired sensations when therapy turns off for approx. 10-15 seconds and then back on, which may lead patient to request surgical intervention for replacement or revision.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Boston Scientific Neuromodulation Corporation
- Distributor
- Worldwide Distribution: US (nationwide) including states of: TN, PA, WA, SC, MN, OH, NC, AK, NY, IL, CA, NJ, KY, IA, MO, UT, FL, IN, LA, AL, MS, MD, WI, GA, VA, TX, AZ, KS, OR, CO, CT, ID, AR, OK, MA, MI, ND, NE, SD, WV, NV, HI, NH, DC, MT, ME, VT, WY, DE, NM. OUS (foreign) countries of: Canada, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Ireland, Israel, Italy, Lebanon, Luxembourg, Netherlands, Norway, Poland, Qatar, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, Portugal, Japan, Australia, New Zealand, Singapore, Argentina, Brazil, Colombia, Puerto Rico
- Lot
- REF/UDI-DI(GTIN)/Serial Number Range/Expiration Date Range: M365SC12160/8714729985082/100188 - 758920, M365SC12320/8714729985099/100635 - 759404
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDANalu TD2 Therapy Disc with firmware, REF: 34014-001, and TD2 Patient Kit, REF: 74011, part of the NaluNovember 4, 2026
- FDAINFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model:August 12, 2026
- FDAIntracept Access Instruments, Description/REF: INTRACEPT ACCESS INSTRUMENTS- DRILL/RLV0035, INTRACEPT ACCESSJuly 15, 2026
- FDABoston Scientific Model SC-4318, Clik X Anchor REF: M365SC43180June 17, 2026
- FDABioButton REF BIOST06040 with Software version 3.0.20 The BioButton System is a remote monitoring wearableAugust 21, 2024
- FDAAurora Series, Aurora 7 lighthead; Models: LED7 (including TV models). The Aurora Series is intended toAugust 21, 2024
Questions about this recall
Was WAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320 recalled?
Yes. FDA published recall Z-2617-2024 on August 21, 2024. Reported hazard: Spinal Cord Stimulation (SCS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset, which may lead to transient loss of stimulation; patients may experience undesired sensations when therapy turns off for approx. 10-15 seconds and then back on, which may lead patient to request surgical intervention for replacement or revision.
What should I do if I have this product?
Voluntary: Firm initiated · Letter