Aprima Access Nonvascular Introducer Set Accessories, Catheter Intended for single-puncture percutaneous access to facilitate placement of an .038 inch (0.97 mm) diameter working wire guide for interventional radiology procedures
Published August 31, 2016 · Recall initiated: April 15, 2016
Summary
FDA records medical device Class II recall Z-2620-2016, published August 31, 2016 and initiated April 15, 2016. Recalling firm: Cook Inc.; Status: Terminated; Reported quantity: 148,334; Lot: All Lots. Catalog prefixes / suffixes NPAS-/-SST Catalog/product number; Global/order number NPAS-100-HC-NT-U-SST G52963 NPAS-100-HC-U-SST G52962 NPAS-101-HC-NT-U-SST G23067 NPAS-101-HC-U-SST G23066 NPAS-104-HC-NT-JWG-U-SST G52966 NPAS-104-HC-NT-U-SST G52965 NPAS-104-HC-U-SST G52964 NPAS-105-HC-NT-U-SST G52967 NPAS-108-HC-NT-U-SST G25516 NPAS-120-HC-NT-U-SST G25517 NSSW-4.0-18-NPAS-100-HC-SST G34253. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cook Inc.
- Distributor
- Worldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, "Palestinian Territory, Occupied", Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, U.S. Virgin Islands, Yugoslavia.
- Lot
- All Lots. Catalog prefixes / suffixes NPAS-/-SST Catalog/product number; Global/order number NPAS-100-HC-NT-U-SST G52963 NPAS-100-HC-U-SST G52962 NPAS-101-HC-NT-U-SST G23067 NPAS-101-HC-U-SST G23066 NPAS-104-HC-NT-JWG-U-SST G52966 NPAS-104-HC-NT-U-SST G52965 NPAS-104-HC-U-SST G52964 NPAS-105-HC-NT-U-SST G52967 NPAS-108-HC-NT-U-SST G25516 NPAS-120-HC-NT-U-SST G25517 NSSW-4.0-18-NPAS-100-HC-SST G34253
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFlexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 TheNovember 10, 2021
- FDAFlexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL0-HC, Order Number G56233 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL0-HC, Order Number G56231 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL1-HC, Order Number G56236 Flexor Introducers andDecember 30, 2020
- FDAShuttle Select Slip-Cath Catheter, Intravascular, Diagnostic The catheters are intended for use inAugust 31, 2016
- FDASOMATOM Definition AS with software version VA48A-SP2; Model # 10430603, computed tomography x-ray systemAugust 31, 2016
Questions about this recall
Was Aprima Access Nonvascular Introducer Set Accessories, Catheter Intended for single-puncture percutaneous access to facilitate placement of an .038 inch (0.97 mm) diameter working wire guide for interventional radiology procedures recalled?
Yes. FDA published recall Z-2620-2016 on August 31, 2016. Reported hazard: Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.
What should I do if I have this product?
Voluntary: Firm initiated · Letter