SOMATOM Definition AS with software version VA48A-SP2; Model # 10430603, computed tomography x-ray system
Published August 31, 2016 · Recall initiated: July 13, 2016
Summary
FDA records medical device Class II recall Z-2624-2016, published August 31, 2016 and initiated July 13, 2016. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 1 unit; Lot: Model # 10430603 Serial # 74192. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to an internal communication error between the firmware and the software of the components, the planned CT scan executes properly, but the injector is not started. Therefore, the contrast agent is not injected and the desired examination result is not achieved. This error only affects the automatic mode or coupled mode and does not affect the manual control of the injector.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- Distributed to: MI, NY, CA, KY, ND, NE
- Lot
- Model # 10430603 Serial # 74192
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono biplane. Model Number: 11327600.July 8, 2026
- FDAAprima Access Nonvascular Introducer Set Accessories, Catheter Intended for single-puncture percutaneousAugust 31, 2016
- FDARESPONSIVE, KNEE: FEMORAL total knee replacement COMPONENT CR/CS, SIZE 5R, REF 90-SRK-311500, QTY: 1, STERILEAugust 31, 2016
Questions about this recall
Was SOMATOM Definition AS with software version VA48A-SP2; Model # 10430603, computed tomography x-ray system recalled?
Yes. FDA published recall Z-2624-2016 on August 31, 2016. Reported hazard: Due to an internal communication error between the firmware and the software of the components, the planned CT scan executes properly, but the injector is not started. Therefore, the contrast agent is not injected and the desired examination result is not achieved. This error only affects the automatic mode or coupled mode and does not affect the manual control of the injector.
What should I do if I have this product?
Voluntary: Firm initiated · Letter