SOMATOM Definition AS with software version VA48A-SP2; Model # 10430603, computed tomography x-ray system

Published August 31, 2016 · Recall initiated: July 13, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to an internal communication error between the firmware and the software of the components, the planned CT scan executes properly, but the injector is not started. Therefore, the contrast agent is not injected and the desired examination result is not achieved. This error only affects the automatic mode or coupled mode and does not affect the manual control of the injector.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
Distributed to: MI, NY, CA, KY, ND, NE
Lot
Model # 10430603 Serial # 74192
Risk classification
Class II
Area (FDA)
device
Country
United States