SOMATOM Definition AS with software version VA48A-SP2; Model # 10430603, computed tomography x-ray system

Published August 31, 2016 · Recall initiated: July 13, 2016

Summary

FDA records medical device Class II recall Z-2624-2016, published August 31, 2016 and initiated July 13, 2016. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 1 unit; Lot: Model # 10430603 Serial # 74192. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to an internal communication error between the firmware and the software of the components, the planned CT scan executes properly, but the injector is not started. Therefore, the contrast agent is not injected and the desired examination result is not achieved. This error only affects the automatic mode or coupled mode and does not affect the manual control of the injector.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
Distributed to: MI, NY, CA, KY, ND, NE
Lot
Model # 10430603 Serial # 74192
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was SOMATOM Definition AS with software version VA48A-SP2; Model # 10430603, computed tomography x-ray system recalled?

Yes. FDA published recall Z-2624-2016 on August 31, 2016. Reported hazard: Due to an internal communication error between the firmware and the software of the components, the planned CT scan executes properly, but the injector is not started. Therefore, the contrast agent is not injected and the desired examination result is not achieved. This error only affects the automatic mode or coupled mode and does not affect the manual control of the injector.