ROTH MOBELI Grab Handles used with EOS System: MOBELI Stabi Vario Grab Handle, ROTH Catalogue Number 14002 96 S, EOS Imaging Catalogue Number 9000111773; and MOBELI Dual Grip Grab Handle, ROTH Catalogue Number 14002 25 S, EOS Imaging Catalogue Number 9000111774.

Published July 5, 2017 · Recall initiated: May 30, 2017

Summary

FDA records medical device Class II recall Z-2626-2017, published July 5, 2017 and initiated May 30, 2017. Recalling firm: EOS Imaging; Status: Terminated; Reported quantity: 8 units each; Lot: ROTH Catalogue Numbers 14002 96 S and 14002 25 S only,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

ROTH MOBELI grab handles could detach and patient could fall if the EOS system operator does not follow instructions. Before each use, the operator must examine the vacuum safety indicator to check the vacuum is monitored and sufficient and the grab handle correctly attached. Do not use grab handles to help patient up. The handles are only for stabilizing the patients hand or arm in the EOS cabin.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
EOS Imaging
Distributor
US Distribution to the states of : Pennsylvania, Massachusetts, Minnesota, Alabama, Michigan, Texas and Washington.
Lot
ROTH Catalogue Numbers 14002 96 S and 14002 25 S only,
Risk classification
Class II
Area (FDA)
device
Country
France

Related recalls

Questions about this recall

Was ROTH MOBELI Grab Handles used with EOS System: MOBELI Stabi Vario Grab Handle, ROTH Catalogue Number 14002 96 S, EOS Imaging Catalogue Number 9000111773; and MOBELI Dual Grip Grab Handle, ROTH Catalogue Number 14002 25 S, EOS Imaging Catalogue Number 9000111774. recalled?

Yes. FDA published recall Z-2626-2017 on July 5, 2017. Reported hazard: ROTH MOBELI grab handles could detach and patient could fall if the EOS system operator does not follow instructions. Before each use, the operator must examine the vacuum safety indicator to check the vacuum is monitored and sufficient and the grab handle correctly attached. Do not use grab handles to help patient up. The handles are only for stabilizing the patients hand or arm in the EOS cabin.

What should I do if I have this product?

Voluntary: Firm initiated · Letter