DIASTAT(R) ANA (anti-Nuclear antibody) EURO DIAGNOSTICA, The DIASTAT(R) anti-nuclear antibody (ANA) test is a qualitative enzyme-linked immunosorbent assay (ELISA) for the detection of ANAs in human serum or EDTA plasma. It detects ANAs against Sm, Sm/RNP, Ro (SS-A), La (SS-B), Scl-70, Jo-1, dsDNA, histone and centromere antigens.
Published July 5, 2017 · Recall initiated: May 22, 2017
Summary
FDA records medical device Class III recall Z-2627-2017, published July 5, 2017 and initiated May 22, 2017. Recalling firm: Euro Diagnostica AB; Status: Terminated; Reported quantity: 501 units; Lot: TS 3577. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
FANA200 kit lot TS 3577 was re-worked due to shortened shelf-life. The alert sticker on the kit box of the lot informed customers that the shelf-life had been shortened and referenced a letter sent to customers for more information, where the new expiration date was stated. The original kit box label remained on the kit box stating the original expiration date, the packaging did not display the correct expiration date.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Euro Diagnostica AB
- Distributor
- MN, TX, NE, PA, Spain, Egypt, Pakistan, Nepal, Canada
- Lot
- TS 3577
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- Sweden
Related recalls
- FDAImmunoscan CCPlus test kit is an enzyme-linked immunosorbent assay (ELISA) for qualitative andAugust 29, 2018
- FDAEuro Diagnostica DIASTAT ANA (anti-nuclear antibody) ELISA DIASTAT ANA (anti-nuclear antibody) A positive ANAJanuary 10, 2018
- FDAROTH MOBELI Grab Handles used with EOS System: MOBELI Stabi Vario Grab Handle, ROTH Catalogue Number 14002 96July 5, 2017
- FDAda Vinci Xi Surgical System Product Usage: The da Vinci Xi Surgical System (IS4000) is designed to assist aJuly 5, 2017
- FDALEGION(R) FEM CONE IMPACTOR HEADS: a) REF 71441976, ID 18MM b) REF 71441977, ID 20MM c) REF 71441978, ID 22MMJuly 5, 2017
- FDAPointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range Standard Set 5July 5, 2017
Questions about this recall
Was DIASTAT(R) ANA (anti-Nuclear antibody) EURO DIAGNOSTICA, The DIASTAT(R) anti-nuclear antibody (ANA) test is a qualitative enzyme-linked immunosorbent assay (ELISA) for the detection of ANAs in human serum or EDTA plasma. It detects ANAs against Sm, Sm/RNP, Ro (SS-A), La (SS-B), Scl-70, Jo-1, dsDNA, histone and centromere antigens. recalled?
Yes. FDA published recall Z-2627-2017 on July 5, 2017. Reported hazard: FANA200 kit lot TS 3577 was re-worked due to shortened shelf-life. The alert sticker on the kit box of the lot informed customers that the shelf-life had been shortened and referenced a letter sent to customers for more information, where the new expiration date was stated. The original kit box label remained on the kit box stating the original expiration date, the packaging did not display the correct expiration date.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail