ADVIA Chemistry A1c_3 CAL (Calibrator) for use on the ADVIA Chemistry Systems, REF/Catalog Number/Siemens Material Number (SMN) 10491408, IVD ADVIA Chemistry A1c_3 Calibrator Information: For in vitro diagnostic use in the calibration of A1c_3 (A1c_3 and A1c_3M) and Total Hemoglobin_3 (tHb_3 and tHb_3M) methods on ADVIA Chemistry systems. The ADVIA Chemistry systems are automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine in random access and batch modes for in vitro diagnostic use.
Published September 24, 2014 · Recall initiated: August 6, 2014
Summary
FDA records medical device Class II recall Z-2643-2014, published September 24, 2014 and initiated August 6, 2014. Recalling firm: Siemens Healthcare Diagnostics; Status: Terminated; Reported quantity: Domestic: 1122 kits; Foreign: 236 kits; Lot: Lots 3HD044 (exp. 11/01/2015), and 3LD068 (exp. 01/01/2016). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The ADVIA Chemistry Systems A1c_3 method may exhibit a percent bias of -9% to +11% for hemoglobin (Hb)A1c samples when using A1c_3 Calibrator lots 3HD044 and 3LD068. Depending on quality control limits, this issue may not have been detected. NOTE: The percent bias range is a percentage of the HbA1c result. It is NOT absolute HbA1c units.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Austria, Brazil, Canada, China, Colombia, Croatia, Cyprus, Denmark, Estonia, Finland, France, Germany, Greece, Guatemala, India, Indonesia, Istanbul, Italy, Lithuania, Mexico, Paraguay, Peru, Poland, Romania, Russian Federation, Serbia, Singapore, Slovakia, South Korea, Spain, Switzerland, Thailand, Uruguay, Vietnam, and Zambia.
- Lot
- Lots 3HD044 (exp. 11/01/2015), and 3LD068 (exp. 01/01/2016)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIMMULITE /IMMULITE 1000 Systems COR Cortisol, REF/Catalog Number LKC01, SMN 10381388, IVD. For in vitroOctober 29, 2014
- FDAADVIA Centaur Immunoassay System (including refurbished units), This system is intended for professional useOctober 8, 2014
- FDAADVIA Centaur XP Immunoassay System (including refurbished units), This system is intended for professionalOctober 8, 2014
- FDAADVIA Centaur XP Immunoassay System, Catalog Numbers/Siemens Material Numbers (SMN) 10285219, 10316507,September 24, 2014
- FDAVac-Assist Suction Units (SUA01 Suction Aspirator Unit w/800 cc Canister). The suction unit is intended forSeptember 24, 2014
- FDAJBAIDS Instrument System Software version 3.5.0.72. The JBAIDS instrument is a portable thermocycler andSeptember 24, 2014
Questions about this recall
Was ADVIA Chemistry A1c_3 CAL (Calibrator) for use on the ADVIA Chemistry Systems, REF/Catalog Number/Siemens Material Number (SMN) 10491408, IVD ADVIA Chemistry A1c_3 Calibrator Information: For in vitro diagnostic use in the calibration of A1c_3 (A1c_3 and A1c_3M) and Total Hemoglobin_3 (tHb_3 and tHb_3M) methods on ADVIA Chemistry systems. The ADVIA Chemistry systems are automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine in random access and batch modes for in vitro diagnostic use. recalled?
Yes. FDA published recall Z-2643-2014 on September 24, 2014. Reported hazard: The ADVIA Chemistry Systems A1c_3 method may exhibit a percent bias of -9% to +11% for hemoglobin (Hb)A1c samples when using A1c_3 Calibrator lots 3HD044 and 3LD068. Depending on quality control limits, this issue may not have been detected. NOTE: The percent bias range is a percentage of the HbA1c result. It is NOT absolute HbA1c units.
What should I do if I have this product?
Voluntary: Firm initiated · Letter