Brilliance BigBore Radiology CT Model number 728244 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
Published August 31, 2016 · Recall initiated: July 20, 2016
Summary
FDA records medical device Class II recall Z-2654-2016, published August 31, 2016 and initiated July 20, 2016. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 24; Lot: 728244 Brilliance BigBore Radiology CT 7375 7781 7614 7633 7708 7789 7721 7760 7252 7380 7390 7514 7052 7244 7384 7533 7172 7200 7546 7427 7452 7407 7453 7586. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
- Lot
- 728244 Brilliance BigBore Radiology CT 7375 7781 7614 7633 7708 7789 7721 7760 7252 7380 7390 7514 7052 7244 7384 7533 7172 7200 7546 7427 7452 7407 7453 7586
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.September 21, 2022
- FDABrilliance 64 Model # 728231, computed tomography x-ray systemAugust 14, 2019
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray systemAugust 14, 2019
- FDABrightView XCT , Model Number 882482July 31, 2019
- FDABrilliance CT Big Bore Oncology, Model number 728243 Product Usage: Computed Tomography X-ray systems intendedAugust 31, 2016
- FDABrilliance CT 16 Power, Model number 728240 Product Usage: Computed Tomography X-ray systems intended toAugust 31, 2016
Questions about this recall
Was Brilliance BigBore Radiology CT Model number 728244 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. recalled?
Yes. FDA published recall Z-2654-2016 on August 31, 2016. Reported hazard: After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail