FDAFDA · Food and drugs·Recall number Z-2654-2016
Brilliance BigBore Radiology CT Model number 728244 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
Published August 31, 2016 · Recall initiated: July 20, 2016
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
- Lot
- 728244 Brilliance BigBore Radiology CT 7375 7781 7614 7633 7708 7789 7721 7760 7252 7380 7390 7514 7052 7244 7384 7533 7172 7200 7546 7427 7452 7407 7453 7586
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States