ADVIA Centaur Immunoassay System, Catalog Numbers/Siemens Material Numbers (SMN) 10284980, 10286140, 10309524, 10309525, 10310210, 10313282, 10314322, 10316248, 10316372, 10316968, 10317060, 10317403, 10319111, 10319433, 10320929, 10321568, 10322149, 10322731, 10323204, 10325015, 10326217, 10327008, 10327379, 10328250, 10328647, 10329364, 10330873, 10331013, 10332617, 10334139, 10334759, 10337512, 10337526, 10339677, 10340551, 10340737, 10341051, 10341110, 10361010, 10361011, 10361012 --- For in vitro diagnostic use This immunoassay system is intended for professional use in a laboratory environment only. Tests performed using this system are intended for in vitro diagnostic use.

Published September 24, 2014 · Recall initiated: August 13, 2014

Summary

FDA records medical device Class II recall Z-2662-2014, published September 24, 2014 and initiated August 13, 2014. Recalling firm: Siemens Healthcare Diagnostics; Status: Terminated; Reported quantity: 10,515 (Domestic: 2771; Foreign: 7744) in total; Lot: All Serial Numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The firm is conducting a field correction for the ADVIA Centaur and ADVIA Centaur XP Immunoassay systems, regarding manually loaded sample racks that are improperly loaded or pushed too far on the sample entry queue during normal operation of the ADVIA Centaur and ADVIA Centaur XP Immunoassay systems. Pushing the sample racks may cause misreads of Sample IDs (SID) if the sample rack loading instructions are not followed exactly as described in the ADVIA Centaur or ADVIA Centaur XP Operators Guide.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Healthcare Diagnostics
Distributor
Worldwide Distribution: US (nationwide) and countries of: Argentina, Austria, Australia, Bahrain, Bangladesh, Belgium, Belgrade, Brazil, Botswana, Canada, Chile, China, Colombia, Croatia, Czech Republic, Cyprus, Denmark, Egypt, Estonia, Finland, France, Germany, Georgia, Great Britain, Hungary, Indonesia, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Kazakstan, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Malta, Martinique, Mexico, Myanmar, New Caledonia, Netherlands, Norway, New Zealand, Pakistan, French Polynesia, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, Slovakia, South Africa, Spain, Switzerland, Thailand, Turkey, United Arab Emirates, Vatican, Venezuela, Vietnam, Yemen, and Zimbabwe.
Lot
All Serial Numbers
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ADVIA Centaur Immunoassay System, Catalog Numbers/Siemens Material Numbers (SMN) 10284980, 10286140, 10309524, 10309525, 10310210, 10313282, 10314322, 10316248, 10316372, 10316968, 10317060, 10317403, 10319111, 10319433, 10320929, 10321568, 10322149, 10322731, 10323204, 10325015, 10326217, 10327008, 10327379, 10328250, 10328647, 10329364, 10330873, 10331013, 10332617, 10334139, 10334759, 10337512, 10337526, 10339677, 10340551, 10340737, 10341051, 10341110, 10361010, 10361011, 10361012 --- For in vitro diagnostic use This immunoassay system is intended for professional use in a laboratory environment only. Tests performed using this system are intended for in vitro diagnostic use. recalled?

Yes. FDA published recall Z-2662-2014 on September 24, 2014. Reported hazard: The firm is conducting a field correction for the ADVIA Centaur and ADVIA Centaur XP Immunoassay systems, regarding manually loaded sample racks that are improperly loaded or pushed too far on the sample entry queue during normal operation of the ADVIA Centaur and ADVIA Centaur XP Immunoassay systems. Pushing the sample racks may cause misreads of Sample IDs (SID) if the sample rack loading instructions are not followed exactly as described in the ADVIA Centaur or ADVIA Centaur XP Operators Guide.

What should I do if I have this product?

Voluntary: Firm initiated · Letter