STAPLER,IS4000; Model number 470298; General and Plastic Surgery: The Intuitive Surgical da Vinci EndoWrist Stapler 45, Stapler 45 Reloads and other Stapler Accessories are intended to be used with the da Vinci Xi Surgical System (Model IS4000) for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device can be used with staple line or tissue butressing material (natural or synthetic).

Published September 7, 2016 · Recall initiated: July 14, 2016

Summary

FDA records medical device Class II recall Z-2675-2016, published September 7, 2016 and initiated July 14, 2016. Recalling firm: Intuitive Surgical, Inc.; Status: Terminated; Reported quantity: 875 instruments (827 a la carte + 48 in kits) total, all models; Lot: All model 470298. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Field failures were reported related to the da Vinci Xi Surgical System EndoWrist Stapler 45 and 30 instruments having an interruption in shaft rotation and / or instrument jaws not opening.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Intuitive Surgical, Inc.
Distributor
Worldwide Distribution - US (nationwide) Internationally to Australia, Belgium, Denmark, France, Germany. Italy. Japan, Spain, Sweden, Switzerland, Taiwan, and Turkey,
Lot
All model 470298
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was STAPLER,IS4000; Model number 470298; General and Plastic Surgery: The Intuitive Surgical da Vinci EndoWrist Stapler 45, Stapler 45 Reloads and other Stapler Accessories are intended to be used with the da Vinci Xi Surgical System (Model IS4000) for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device can be used with staple line or tissue butressing material (natural or synthetic). recalled?

Yes. FDA published recall Z-2675-2016 on September 7, 2016. Reported hazard: Field failures were reported related to the da Vinci Xi Surgical System EndoWrist Stapler 45 and 30 instruments having an interruption in shaft rotation and / or instrument jaws not opening.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail