Back to search
FDAFDA · Food and drugs·Recall number Z-2676-2020

Ablation Galil Technology, IcePearl 2.1 CX Needle Cryoablation Needle, Prescription Only Galil Medicals 2.1 CX Cryoablation Needles are intended for cryoablative destruction of tissue during surgical procedures.

Published July 29, 2020 · Recall initiated: June 4, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The needle surface deteriorated due to unexpected electrolysis during procedure.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Boston Scientific Corporation
Distributor
Nationwide Distribution to states of: AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, HI, IA, ID, KS, LA, MA, MD, ME, MN, MO, MS, NC, ND, NE, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI WV, and WY and the country of: Canada.
Lot
UPN: FPRPR3603, GTIN: 7290104831010, Batch: A1564, A1521, A6723, A6908, A6929, A7015, T0025, T0056, T0137, T0138, T0269, U0041
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card