Back to search
FDAFDA · Food and drugs·Recall number Z-2679-2020

Ablation Galil Technology, IcePearl®2.1 CX Prostate®Kit Visual Ice System, Prescription Only Galil Medicals 2.1 CX Cryoablation Needles are intended for cryoablative destruction of tissue during surgical procedures.

Published July 29, 2020 · Recall initiated: June 4, 2020

Summary

FDA records medical device Class II recall Z-2679-2020, published July 29, 2020 and initiated June 4, 2020. Recalling firm: Boston Scientific Corporation; Status: Terminated; Reported quantity: 193 devices; Lot: UPN:FPRPR3606, GTIN: 7290104831041, Batch: A1148, A1171, A1248, A1460, A1561, A1582, A1936, A2121, A6702, A6743, A6856, A6925, A7013, A7026, A7088, A7148, T0054, T0136, T0141, T0224, T0236, T0237, T0268, T0437, T0592, U0082, U0027, U0093, U0226, U0227. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The needle surface deteriorated due to unexpected electrolysis during procedure.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Boston Scientific Corporation
Distributor
Nationwide Distribution to states of: AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, HI, IA, ID, KS, LA, MA, MD, ME, MN, MO, MS, NC, ND, NE, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI WV, and WY and the country of: Canada.
Lot
UPN:FPRPR3606, GTIN: 7290104831041, Batch: A1148, A1171, A1248, A1460, A1561, A1582, A1936, A2121, A6702, A6743, A6856, A6925, A7013, A7026, A7088, A7148, T0054, T0136, T0141, T0224, T0236, T0237, T0268, T0437, T0592, U0082, U0027, U0093, U0226, U0227
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Ablation Galil Technology, IcePearl®2.1 CX Prostate®Kit Visual Ice System, Prescription Only Galil Medicals 2.1 CX Cryoablation Needles are intended for cryoablative destruction of tissue during surgical procedures. recalled?

Yes. FDA published recall Z-2679-2020 on July 29, 2020. Reported hazard: The needle surface deteriorated due to unexpected electrolysis during procedure.

What should I do if I have this product?

Voluntary: Firm initiated · Letter