Multiphor II Electrophoresis System GE HealthCare Product Number: 18-1018-06 PerkinElmer Part Number: 2117-004 Product Usage: Multiphor II Electrophoresis System is a versatile modular electrophoresis system for horizontal electrophoresis, isoelectric focusing, 2-D electrophoresis and electrophoretic transfer Note: The Multiphor II Electrophoresis System (Class I) is the platform used with the hemoglobin electrophoresis gel kit ( Class II) sold by PerkinElmer as the Resolve Hemoglobin Kit (K050709: Product code JJN). The RESOLVE Systems Hemoglobin kit (K050709) is designed to separate whole blood, cord blood or dried blood spot specimen for detection of normal and variant hemoglobins by isoelectric focusing.
Published July 12, 2017 · Recall initiated: May 15, 2017
Summary
FDA records medical device Class II recall Z-2685-2017, published July 12, 2017 and initiated May 15, 2017. Recalling firm: Perkinelmer Life Sciences, Inc.; Status: Terminated; Reported quantity: 241 instruments; Lot: Serial Number: 1171221 1171231 1171232 1171234 1181923 1181925 1181928 1187499 1189475 1189484 1195060 1195074 1197741 1197743 1197752 1197753 1200106 1202993 1202994 1251991 1252000 1253298 1253301 1253307 1254797 1254798 1255972 1264957 1264960 1302820 1302823 1313213 1320921 1320925 1320927 1331757 1383503 1383506 1406432 1413361 1434395 1434396 1434399 1435683 1436876 1442735 1463140 1465792 1465793 1465794 1480331 1480829 1480834 1486179 1486180 1487157 1487158 1487159 1487160 1487161 1487162 1487182 1497464 1497490 1509220 1525734 1525737 1525739 1525741 1525744 1525745 1525748 1540014 1540016 1540018 1540019 1540020 1550838 1550839 1561681 1564304 1604476 1604481 1607341 1607345 1607346 1624871 1624873. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for electric shock in the case of a failure to install the grommets, supplied with the system.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Perkinelmer Life Sciences, Inc.
- Distributor
- US Nationwide Distribution
- Lot
- Serial Number: 1171221 1171231 1171232 1171234 1181923 1181925 1181928 1187499 1189475 1189484 1195060 1195074 1197741 1197743 1197752 1197753 1200106 1202993 1202994 1251991 1252000 1253298 1253301 1253307 1254797 1254798 1255972 1264957 1264960 1302820 1302823 1313213 1320921 1320925 1320927 1331757 1383503 1383506 1406432 1413361 1434395 1434396 1434399 1435683 1436876 1442735 1463140 1465792 1465793 1465794 1480331 1480829 1480834 1486179 1486180 1487157 1487158 1487159 1487160 1487161 1487162 1487182 1497464 1497490 1509220 1525734 1525737 1525739 1525741 1525744 1525745 1525748 1540014 1540016 1540018 1540019 1540020 1550838 1550839 1561681 1564304 1604476 1604481 1607341 1607345 1607346 1624871 1624873
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAQUELICIN (Succinylcholine Chloride) Injection, USP 20 mg/mL in a) 5 mL vial (NDC 15082-814-67), b) 7mL vialJuly 19, 2017
- FDAProteus 235, Proton Therapy System Product Usage: The Proton Therapy System - Proteus 235 is a medical deviceJuly 12, 2017
- FDATenaxis Medical ArterX(TM) Surgical Sealant, (a) 4mL, REF 9002, and (b) 5mL REF 9006, STERILE R, Store at 2oCJuly 12, 2017
- FDABoston Scientific Extractor Pro RX-S Retrieval Balloon Catheter 12-15mm Below UPN: M00547340July 12, 2017
- FDACoulter PrepPlus, Part Number 286600, Product Usage: The COULTER PrepPlus is a microprocessor-controlledJuly 12, 2017
- FDAJelco, Hypodermic needle Pro® needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SBJuly 12, 2017
Questions about this recall
Was Multiphor II Electrophoresis System GE HealthCare Product Number: 18-1018-06 PerkinElmer Part Number: 2117-004 Product Usage: Multiphor II Electrophoresis System is a versatile modular electrophoresis system for horizontal electrophoresis, isoelectric focusing, 2-D electrophoresis and electrophoretic transfer Note: The Multiphor II Electrophoresis System (Class I) is the platform used with the hemoglobin electrophoresis gel kit ( Class II) sold by PerkinElmer as the Resolve Hemoglobin Kit (K050709: Product code JJN). The RESOLVE Systems Hemoglobin kit (K050709) is designed to separate whole blood, cord blood or dried blood spot specimen for detection of normal and variant hemoglobins by isoelectric focusing. recalled?
Yes. FDA published recall Z-2685-2017 on July 12, 2017. Reported hazard: Potential for electric shock in the case of a failure to install the grommets, supplied with the system.
What should I do if I have this product?
Voluntary: Firm initiated · Letter