Brilliance CT System 6/10/16/16P/40 configuration. No packaging, sold as single unit. Computed tomography scanner intended to provide cross sectional images of the human body and visualization of the internal organs.
Published October 1, 2014 · Recall initiated: December 27, 2005
Summary
FDA records medical device Class II recall Z-2709-2014, published October 1, 2014 and initiated December 27, 2005. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 364; Lot: 728256, 728251, 728246, 728240, and 728235. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The system has a gap that can appear between the upper and lower tilt stand covers. The fingers may be squeezed if a person was to place them in this region while the gantry was tilting from an angle to zero tilt angle.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- Worldwide Distribution -- AZ, CO, GA, IA, IL, IN, KS, KY, MD, MI, MN, MS, NC, NY, OH, OK, TX, VT, and WI; and, countries of Singapore, New Zealand, India, Australia, Korea, Japan, Korea, China, India, Taiwan, Australia, Turkey, France, South Africa, Spain, Italy, Austria, Belgium, Switzerland, Israel, Netherlands, England, and Brazil.
- Lot
- 728256, 728251, 728246, 728240, and 728235.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.September 21, 2022
- FDABrilliance 64 Model # 728231, computed tomography x-ray systemAugust 14, 2019
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray systemAugust 14, 2019
- FDABrightView XCT , Model Number 882482July 31, 2019
- FDACUBICIN (daptomycin for injection), 500 mg single-use vial, Rx only, Manufactured for: Cubist Pharmaceuticals,October 8, 2014
- FDADako Autostainer Link 48 with software version Dako Link 4.0.3, Model No. AS480 Product Usage: For in vitroOctober 1, 2014
Questions about this recall
Was Brilliance CT System 6/10/16/16P/40 configuration. No packaging, sold as single unit. Computed tomography scanner intended to provide cross sectional images of the human body and visualization of the internal organs. recalled?
Yes. FDA published recall Z-2709-2014 on October 1, 2014. Reported hazard: The system has a gap that can appear between the upper and lower tilt stand covers. The fingers may be squeezed if a person was to place them in this region while the gantry was tilting from an angle to zero tilt angle.
What should I do if I have this product?
Voluntary: Firm initiated · Letter