NOMAD Pro Packaged X-Ray System, Part Number 08500009; NOMAD Pro Vet Packaged X-Ray System, Part Number 08500011; NOMAD Pro Packaged X-Ray System - Europe, Part Number 08500013; NOMAD Pro X-Ray System, Remote Config., Part Number 08500017; NOMAD Pro Packaged X-Ray System JP, Part Number 08500019
Published September 14, 2016 · Recall initiated: July 18, 2016
Summary
FDA records medical device Class II recall Z-2716-2016, published September 14, 2016 and initiated July 18, 2016. Recalling firm: Aribex Inc; Status: Terminated; Reported quantity: 16,871; Lot: Part Number 08500009 - Serial Number Range BX-011008-01 - XB-141219049; Part Number 08500011- Serial Number Range XC -090305-03 - XC-140312-02; Part Number 08500013 - Serial Number Range XB-110411-01 - XB-140616-13; Part Number 08500017 - Serial Number Range XB-120911-04 - XB-141210-50; Part Number 08500019 - Serial Number Range XB-120517-09 - XB-121212-24. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Aribex Inc
- Distributor
- AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA. HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY , Puerto Rico, Afghanistan, Argentina, Australia, Bahamas, Bahrain, Belgium, Bermuda, Chile, China, Columbia, Denmark, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kuwait, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Saudi Arabia, Papua New Guinea, Philippines, Poland, Qatar, Saudi Arabia, Slovenia, South Africa, Sweden, Switzerland, Thailand, Trinidad and Tobago, Turkey, UAE, United Kingdom, Venezuela, Vietnam
- Lot
- Part Number 08500009 - Serial Number Range BX-011008-01 - XB-141219049; Part Number 08500011- Serial Number Range XC -090305-03 - XC-140312-02; Part Number 08500013 - Serial Number Range XB-110411-01 - XB-140616-13; Part Number 08500017 - Serial Number Range XB-120911-04 - XB-141210-50; Part Number 08500019 - Serial Number Range XB-120517-09 - XB-121212-24
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDANOMAD Pro2 Packaged X-Ray System, Part Number 08500021; NOMAD Pro2 Packaged X-Ray System Black, Part NumberSeptember 14, 2016
- FDAADVANCE(R) II MODULAR TIBIAL BASE W/SCREWHOLES, Knee Tibial Base, REF KTTIHA41, SIZE 4+; SURFACE POROUS, HASeptember 14, 2016
- FDAADVANCE(R) II MODULAR TIBIAL BASE W/SCREWHOLES, Knee Tibial Base, REF KTTIHA30, SIZE 3 STD; SURFACE POROUS, HASeptember 14, 2016
- FDAMedical Linear AcceleratorSeptember 14, 2016
- FDAADVANCE(R) II MODULAR TIBIAL BASE W/SCREWHOLES, Knee Tibial Base, REF KTTIHA10, SIZE 1 STD; SURFACE POROUS, HASeptember 14, 2016
- FDAPlum A+ Hyperbaric Infusion Pump. List Number 11005. The Plum A+ Infusion Systems are a multi-purpose familySeptember 14, 2016
Questions about this recall
Was NOMAD Pro Packaged X-Ray System, Part Number 08500009; NOMAD Pro Vet Packaged X-Ray System, Part Number 08500011; NOMAD Pro Packaged X-Ray System - Europe, Part Number 08500013; NOMAD Pro X-Ray System, Remote Config., Part Number 08500017; NOMAD Pro Packaged X-Ray System JP, Part Number 08500019 recalled?
Yes. FDA published recall Z-2716-2016 on September 14, 2016. Reported hazard: Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.
What should I do if I have this product?
Voluntary: Firm initiated · Letter