NOMAD Pro Packaged X-Ray System, Part Number 08500009; NOMAD Pro Vet Packaged X-Ray System, Part Number 08500011; NOMAD Pro Packaged X-Ray System - Europe, Part Number 08500013; NOMAD Pro X-Ray System, Remote Config., Part Number 08500017; NOMAD Pro Packaged X-Ray System JP, Part Number 08500019

Published September 14, 2016 · Recall initiated: July 18, 2016

Summary

FDA records medical device Class II recall Z-2716-2016, published September 14, 2016 and initiated July 18, 2016. Recalling firm: Aribex Inc; Status: Terminated; Reported quantity: 16,871; Lot: Part Number 08500009 - Serial Number Range BX-011008-01 - XB-141219049; Part Number 08500011- Serial Number Range XC -090305-03 - XC-140312-02; Part Number 08500013 - Serial Number Range XB-110411-01 - XB-140616-13; Part Number 08500017 - Serial Number Range XB-120911-04 - XB-141210-50; Part Number 08500019 - Serial Number Range XB-120517-09 - XB-121212-24. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Aribex Inc
Distributor
AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA. HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY , Puerto Rico, Afghanistan, Argentina, Australia, Bahamas, Bahrain, Belgium, Bermuda, Chile, China, Columbia, Denmark, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kuwait, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Saudi Arabia, Papua New Guinea, Philippines, Poland, Qatar, Saudi Arabia, Slovenia, South Africa, Sweden, Switzerland, Thailand, Trinidad and Tobago, Turkey, UAE, United Kingdom, Venezuela, Vietnam
Lot
Part Number 08500009 - Serial Number Range BX-011008-01 - XB-141219049; Part Number 08500011- Serial Number Range XC -090305-03 - XC-140312-02; Part Number 08500013 - Serial Number Range XB-110411-01 - XB-140616-13; Part Number 08500017 - Serial Number Range XB-120911-04 - XB-141210-50; Part Number 08500019 - Serial Number Range XB-120517-09 - XB-121212-24
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was NOMAD Pro Packaged X-Ray System, Part Number 08500009; NOMAD Pro Vet Packaged X-Ray System, Part Number 08500011; NOMAD Pro Packaged X-Ray System - Europe, Part Number 08500013; NOMAD Pro X-Ray System, Remote Config., Part Number 08500017; NOMAD Pro Packaged X-Ray System JP, Part Number 08500019 recalled?

Yes. FDA published recall Z-2716-2016 on September 14, 2016. Reported hazard: Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.

What should I do if I have this product?

Voluntary: Firm initiated · Letter