NOMAD Pro2 Packaged X-Ray System, Part Number 08500021; NOMAD Pro2 Packaged X-Ray System Black, Part Number 08500022; NOMAD Pro2 Vet Packaged X-Ray System, Part Number 08500023; NOMAD Pro2 Vet Packaged X-Ray System Black, Part Number 08500024; NOMAD Pro2 Packaged X-Ray System, GBL, Part Number 08500025 and Part Number 08500026; NOMAD Pro2 Packaged X-Ray System, GLB, Vet, Part Number 08500027 and Part Number 08500028; NOMAD Pro2 Package X-Ray System, GLB, Black, Part Number 08500029

Published September 14, 2016 · Recall initiated: July 18, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Aribex Inc
Distributor
AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA. HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY , Puerto Rico, Afghanistan, Argentina, Australia, Bahamas, Bahrain, Belgium, Bermuda, Chile, China, Columbia, Denmark, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kuwait, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Saudi Arabia, Papua New Guinea, Philippines, Poland, Qatar, Saudi Arabia, Slovenia, South Africa, Sweden, Switzerland, Thailand, Trinidad and Tobago, Turkey, UAE, United Kingdom, Venezuela, Vietnam
Lot
Part Number 08500021 - Serial Number Range XD-100921-05 - XD-150109-11, 10000 - 20811, XD-140929-39 - XD-150123-20; Part Number 08500022 - Serial Number Range XE-130528-03 - XE-150123-38, 10000 - 20541, XE-140908-32 - XE-150123-40; Part Number 08500023 - Serial Number Range XF-121008-05 - XF-150205-03, 10003 - 20032, XE-150205-01; Part Number 08500024 - Serial Number Range XG-130520-01 - XG-141210-05, 1000 - 20019; Part Number 08500025 - Serial Number Range XI-140303-02 - XI-150210-9, XI-150202-04 - XI-150210-19; Part Number 08500026 - Serial Number Range 10039 - 20041; Part Number 08500027 - Serial Number Range 1000 - 20006, Part Number 08500028 - Serial Number Range 10000 - 20030; Part Number 08500029 - Serial Number Range 10000.