Optima XR220amx, Mobile Digital-Ready Radiographic System
Published July 22, 2026 · Recall initiated: June 5, 2026
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- GE Medical Systems, LLC
- Distributor
- Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.
- Lot
- UDI-DI: 00840682115261; System IDs: 05249MOB02.OLD, 205820DP1, 228818XR220, 229276M1, 240566P1, 301295OPTIMA2, 303788SBOPT1, 314251P5, 352LEEDP1, 386FLGDPXR3, 417347DPORT1, 512SAMCSB, 516632OPTIMA2, 540981ODP7, 563264OPT, 601445DP1, 615338HDP01, 661949DP2, 713441OPCDX1, 717657OPT1, 718206OPT240A, 727848SB1, 732321DP1, 757736DP3, 813302OPT1, 828213DXP1, 828286OPT1, 843847220D, 850070941, 865545DP3, 915821OPT1, 941342OP221, 972566XR220E, 985447TM4, SA1208MM05, SB2201700085WK, XR1034798WK7, XR1037005WK4