Back to search
FDAFDA · Food and drugs·Recall number Z-2733-2015

Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of thyroglobulin levels in human serum and plasma using the Access Immunoassay Systems.

Published September 9, 2015 · Recall initiated: July 28, 2015

Summary

FDA records medical device Class II recall Z-2733-2015, published September 9, 2015 and initiated July 28, 2015. Recalling firm: Beckman Coulter Inc.; Status: Terminated; Reported quantity: 3,035 (U.S.) and 7,123 (International); Lot: Catalogue number: 33860 Lot numbers: 431990 434727 430969 522896 523251 Expiration date: 10/31/16 12/31/16 08/31/16 04/30/17 05/31/17. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Beckman Coulter, Inc. is recalling Access Thyroglobulin reagent packs because some may exhibit greater than 10% imprecision at concentrations <10 ng/mL.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Beckman Coulter Inc.
Distributor
Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Algeria, Australia, Austria, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Korea, Republic of, Lebanon, Lithuania, Mexico, Morocco, New Zealand, Oman, Philippines, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Vietnam and Yemen.
Lot
Catalogue number: 33860 Lot numbers: 431990 434727 430969 522896 523251 Expiration date: 10/31/16 12/31/16 08/31/16 04/30/17 05/31/17
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of thyroglobulin levels in human serum and plasma using the Access Immunoassay Systems. recalled?

Yes. FDA published recall Z-2733-2015 on September 9, 2015. Reported hazard: Beckman Coulter, Inc. is recalling Access Thyroglobulin reagent packs because some may exhibit greater than 10% imprecision at concentrations <10 ng/mL.

What should I do if I have this product?

Voluntary: Firm initiated · Letter