Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.
Published August 5, 2026 · Recall initiated: June 25, 2026
Summary
FDA records medical device Class I recall Z-2735-2026, published August 5, 2026 and initiated June 25, 2026. Recalling firm: ResMed Ltd.; Status: Ongoing; Reported quantity: 168,069 devices globally; Lot: Catalog Number: Astral 100 UDI-DI code: 00619498003259 Serial Numbers: Astral Ventilators-Serial number < 22241890149 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070 Catalog Number: Astral 150 UDI-DI code: 00619498270033 Serial Numbers: Astral Ventilator-Serial numbers < 22241440282 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- ResMed Ltd.
- Distributor
- Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Great Britain, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Islamic Republic of Iran, Israel, Italy, Japan, Jordan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Saint Martin (French part), Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Uruguay, and Viet Nam.
- Lot
- Catalog Number: Astral 100 UDI-DI code: 00619498003259 Serial Numbers: Astral Ventilators-Serial number < 22241890149 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070 Catalog Number: Astral 150 UDI-DI code: 00619498270033 Serial Numbers: Astral Ventilator-Serial numbers < 22241440282 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- Australia
Related recalls
- FDAIngenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity CT UpgradesAugust 5, 2026
- FDAPower Over Fiber Base ModuleAugust 5, 2026
- FDAIQon Spectral CT IQon Spectral CT UpgradesAugust 5, 2026
- FDAPack Angiography NCMC 3/Cs BANNER, REF: NCAG62I, containing Medline Namic Angiographic Star Off HandleAugust 5, 2026
- FDABrilliance iCT Brilliance iCT Upgrades Brilliance iCT SP Brilliance CT 64 Channel CT6000 Brilliance 64August 5, 2026
- FDAPermobil M3 Corpus Power Wheelchair with Heavy Duty Shock REF:120031, Model version 3August 5, 2026
Questions about this recall
Was Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation. recalled?
Yes. FDA published recall Z-2735-2026 on August 5, 2026. Reported hazard: Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.
What should I do if I have this product?
Voluntary: Firm initiated · Letter