Summary
FDA records medical device Class II recall Z-2795-2026, published August 5, 2026 and initiated February 29, 2024. Recalling firm: Permobil Inc; Status: Ongoing; Reported quantity: 140 units; Lot: All Units operating with firmware version 1.26.2/UDI:(01)17330818345674. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
This recall has been reported retroactively. The power wheelchair firmware lacks a speed restriction to maintain stability when the seat lift exceeds 50 mm, which may cause the user to fall from the wheelchair or the wheelchair to tip over,
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Permobil Inc
- Distributor
- Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CAN.
- Lot
- All Units operating with firmware version 1.26.2/UDI:(01)17330818345674
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAM5 Corpus Power Wheelchair REF,120032August 5, 2026
- FDAPack Angiography NCMC 3/Cs BANNER, REF: NCAG62I, containing Medline Namic Angiographic Star Off HandleAugust 5, 2026
- FDAAMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) ENT Pack, Part Number AMS10411B; 2) ENTAugust 5, 2026
- FDAPower Over Fiber Base ModuleAugust 5, 2026
- FDAAMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) ENT Pack, Pack Number AMS12206; 2) ENTAugust 5, 2026
- FDAResmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittentAugust 5, 2026
Questions about this recall
Was Permobil M3 Corpus Power Wheelchair with Heavy Duty Shock REF:120031, Model version 3 recalled?
Yes. FDA published recall Z-2795-2026 on August 5, 2026. Reported hazard: This recall has been reported retroactively. The power wheelchair firmware lacks a speed restriction to maintain stability when the seat lift exceeds 50 mm, which may cause the user to fall from the wheelchair or the wheelchair to tip over,
What should I do if I have this product?
Voluntary: Firm initiated · Letter