FDAFDA · Food and drugs·Recall number Z-2804-2016

The VITROS® Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITROS® Chemistry Products HbA1c Reagent Kit, Generation 6 (GEN 6) and Assay Data Disk (ADD), Data Release Versions (DRV) 5873, 5874, 5875, or 5876, is used to calibrate the VITROS 5,1 FS Chemistry System, VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the determination of percent glycated hemoglobin (HbA1c) in human whole blood.

Published September 21, 2016 · Recall initiated: April 19, 2016

Summary

FDA records medical device Class III recall Z-2804-2016, published September 21, 2016 and initiated April 19, 2016. Recalling firm: Ortho-Clinical Diagnostics; Status: Terminated; Reported quantity: 520 Units; Lot: Lot 3155 Product Codes: 6842906 & 6801876. Class III means the product is not likely to cause adverse health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a successful calibration of the assay.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ortho-Clinical Diagnostics
Distributor
Worldwide Distribution - US Nationwide in the states of: AL, AR, AZ, CA, CT, DC, FL, GA, IL, IN, KY, LA, MD, MI, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, WA, WV & Puerto Rico;** Foreign: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy & Spain.
Lot
Lot 3155 Product Codes: 6842906 & 6801876
Risk classification
Class III
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was The VITROS® Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITROS® Chemistry Products HbA1c Reagent Kit, Generation 6 (GEN 6) and Assay Data Disk (ADD), Data Release Versions (DRV) 5873, 5874, 5875, or 5876, is used to calibrate the VITROS 5,1 FS Chemistry System, VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the determination of percent glycated hemoglobin (HbA1c) in human whole blood. recalled?

Yes. FDA published recall Z-2804-2016 on September 21, 2016. Reported hazard: There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a successful calibration of the assay.

What should I do if I have this product?

Voluntary: Firm initiated · Letter