FDAFDA · Food and drugs·Recall number Z-2805-2026
M5 Corpus Power Wheelchair REF,120032
Published August 5, 2026 · Recall initiated: June 19, 2025
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Permobil Inc
- Distributor
- Worldwide - US: Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CANADA.
- Lot
- Power Platform firmware versions 1.27 through 1.57 UDI-DI 17330818334562
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States