FDAFDA · Food and drugs·Recall number Z-2805-2026

M5 Corpus Power Wheelchair REF,120032

Published August 5, 2026 · Recall initiated: June 19, 2025

Summary

FDA records medical device Class II recall Z-2805-2026, published August 5, 2026 and initiated June 19, 2025. Recalling firm: Permobil Inc; Status: Ongoing; Reported quantity: 715 wheelchairs; Lot: Power Platform firmware versions 1.27 through 1.57 UDI-DI 17330818334562. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Permobil Inc
Distributor
Worldwide - US: Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CANADA.
Lot
Power Platform firmware versions 1.27 through 1.57 UDI-DI 17330818334562
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was M5 Corpus Power Wheelchair REF,120032 recalled?

Yes. FDA published recall Z-2805-2026 on August 5, 2026. Reported hazard: This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit