Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Version: 5.7.0 Product Description: The MC2 Portable X-Ray System ( MC2 System or MC2 ) is a portable and handheld x-ray system designed to aid clinicians with point of care visualization through diagnostic x-rays of extremities and shoulders. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The MC2 consists of two major system components: the emitter and the cassette. The MC2 emitter and cassette are battery-powered and can be charged with either a wired or wireless charger. The system is intended to interface wirelessly to an external tablet or display when used with the OXOS Device App. The MC2 utilizes an infrared Tracking System and Viewfinder that allows the operator to position the emitter within a safe source-to-image distance (SID) and allows the operator to visualize the anatomy and x-ray field projection prior to initiating an x-ray acquisition. The MC2 also utilizes a LIDAR system to ensure patient safety by maintaining a safe source-to-skin distance. Component: The MC2 Portable X-Ray System is not a component.
Published August 12, 2026 · Recall initiated: June 11, 2026
Summary
FDA records medical device Class II recall Z-2857-2026, published August 12, 2026 and initiated June 11, 2026. Recalling firm: Oxos Medical Inc; Status: Ongoing; Reported quantity: 77; Lot: Lot Code: MC2; UDI 00860008892782; Serial Numbers 1523, 1527, 1488, 1528, 1529, 1544 MC2.1; UDI 00850079606043; Serial Numbers: 1461, 1462, 1547 Full UDI for affected devices: (01)00860008892782(11)260107(21)1523 (01)00860008892782(11)260129(21)1527 (01)00860008892782(11)260109(21)1488 (01)00860008892782(11)260123(21)1528 (01)00860008892782(11)260130(21)1529 (01)00860008892782(11)260227(21)1544 (01)00850079606043(11)260407(21)1461 (01)00850079606043(11)260407(21)1462 (01)00850079606043(11)260408(21)1547. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to a defective circuit board that overheats
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Oxos Medical Inc
- Distributor
- US Nationwide distribution in the states of FL, GA, IN, SC, TX, VA.
- Lot
- Lot Code: MC2; UDI 00860008892782; Serial Numbers 1523, 1527, 1488, 1528, 1529, 1544 MC2.1; UDI 00850079606043; Serial Numbers: 1461, 1462, 1547 Full UDI for affected devices: (01)00860008892782(11)260107(21)1523 (01)00860008892782(11)260129(21)1527 (01)00860008892782(11)260109(21)1488 (01)00860008892782(11)260123(21)1528 (01)00860008892782(11)260130(21)1529 (01)00860008892782(11)260227(21)1544 (01)00850079606043(11)260407(21)1461 (01)00850079606043(11)260407(21)1462 (01)00850079606043(11)260408(21)1547
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAmunevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart GlassesAugust 12, 2026
- FDABrand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence Model/Catalog Number:August 12, 2026
- FDANerivio Infinity. Model/Catalog Number: FGD000100. Software Version: FW Version 4.1.2.0, App version 7.28.0.11August 12, 2026
- FDAAnti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw SystemAugust 12, 2026
- FDABrand Name: WASPLab Product Name: WASPLab Model/Catalog Number: W087 Software Version: WebApp version 2.33.1,August 12, 2026
- FDAFlow Anesthesia Machine Flow-i C20 6677200 Flow-i C30 6677300 Flow-i C40 6677400 Flow-i C20 6888520 Flow-i C30August 12, 2026
Questions about this recall
Was Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Version: 5.7.0 Product Description: The MC2 Portable X-Ray System ( MC2 System or MC2 ) is a portable and handheld x-ray system designed to aid clinicians with point of care visualization through diagnostic x-rays of extremities and shoulders. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The MC2 consists of two major system components: the emitter and the cassette. The MC2 emitter and cassette are battery-powered and can be charged with either a wired or wireless charger. The system is intended to interface wirelessly to an external tablet or display when used with the OXOS Device App. The MC2 utilizes an infrared Tracking System and Viewfinder that allows the operator to position the emitter within a safe source-to-image distance (SID) and allows the operator to visualize the anatomy and x-ray field projection prior to initiating an x-ray acquisition. The MC2 also utilizes a LIDAR system to ensure patient safety by maintaining a safe source-to-skin distance. Component: The MC2 Portable X-Ray System is not a component. recalled?
Yes. FDA published recall Z-2857-2026 on August 12, 2026. Reported hazard: Due to a defective circuit board that overheats
What should I do if I have this product?
Voluntary: Firm initiated · Letter