LifeCare PCA 3, PCA Serial List Number 12384 Allows for clinician administration or self-administration of analgesic medications.

Published September 28, 2016 · Recall initiated: March 8, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Screw rotation error occurring when the pump detects a mismatch between the number of steps given to the motor driving delivery and the number of steps the motor actually took. When this malfunction occurs, the pump will alarm and infusion stops.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Hospira Inc.
Distributor
United States (Nationwide) Distribution including Puerto Rico and American Virgin Islands.
Lot
Serial Number 12921204 12922976 12924008 12925861 12925924 12925998 12928528 12928807 12929016 12929126 12920432 12920820 12920856 12920915 12921404 12921413 12922304 12922976 12925861 12925998 12927215 12929016 12931044 12931141 12931155 12931192 12931244 12931342 12931361 12931413 12931610 12931639 12931649 12931911 12921480 12921487 12923723 12924326 12924350 12924546 12925259 12925447 12926315 12927938 12928075 12928614 12930418 12930709 12930890 12931441 12931760 12931911 12932487 12932717 12920754 12920791 12920897 12921063 12921613 12922062 12923441 12924161 12926859 12927756 12928775 12929346 12929438 12931942 12920112
Risk classification
Class II