GEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 00077415010 00077430010 Product Description: The GEM Premier 7000 with iQM3 is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory.
Published August 12, 2026 · Recall initiated: May 21, 2026
Summary
FDA records medical device Class II recall Z-2899-2026, published August 12, 2026 and initiated May 21, 2026. Recalling firm: Instrumentation Laboratory; Status: Ongoing; Reported quantity: 32 units; Lot: Part No.00077407508 UDI-DI 00195226000764 Lot 260504E; Part No. 00077407510 UDI-DI 00195226000665 Lot 260505B; Part No. 00077415004 UDI-DI 00195226000856 Lot 260505A; Part No. 00077415008 UDI-DI 00195226000757 Lot No.260505A; Part No. 00077415010 UDI-DI 00195226000658 Lot 260505C; Part No. 00077430010 UDI-DI 00195226000641 Lot 260505B. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Instrumentation Laboratory
- Distributor
- Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, KY, MN, MO, MT, OH, PA, VA, WI, WV and the countries of China, Germany, Italy.
- Lot
- Part No.00077407508 UDI-DI 00195226000764 Lot 260504E; Part No. 00077407510 UDI-DI 00195226000665 Lot 260505B; Part No. 00077415004 UDI-DI 00195226000856 Lot 260505A; Part No. 00077415008 UDI-DI 00195226000757 Lot No.260505A; Part No. 00077415010 UDI-DI 00195226000658 Lot 260505C; Part No. 00077430010 UDI-DI 00195226000641 Lot 260505B
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnosticOctober 21, 2026
- FDAHemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)October 21, 2026
- FDAGEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011August 12, 2026
- FDABrand Name: U2 Total Knee System Product Name: Femoral offset bushing, 2mm and 6mm Model/Catalog Number:August 12, 2026
- FDABrand Name: U2 Total Knee System Product Name: Femoral offset bushing, 0mm and 4mm Model/Catalog Number:August 12, 2026
- FDAGEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/CatalogAugust 5, 2026
Questions about this recall
Was GEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 00077415010 00077430010 Product Description: The GEM Premier 7000 with iQM3 is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. recalled?
Yes. FDA published recall Z-2899-2026 on August 12, 2026. Reported hazard: The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
What should I do if I have this product?
Voluntary: Firm initiated · Letter