Back to search
FDAFDA · Food and drugs·Recall number Z-2900-2026

GEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011 00055445010 00055445011 00055360010.A

Published August 12, 2026 · Recall initiated: May 21, 2026

Summary

FDA records medical device Class II recall Z-2900-2026, published August 12, 2026 and initiated May 21, 2026. Recalling firm: Instrumentation Laboratory; Status: Ongoing; Reported quantity: 109 units; Lot: Part No. 00055360010 UDI-DI 08426950807599 Lots 260505K, 260504AG, 260505F; Part No. 00055360011 UDI-DI 08426950807605 Lots 260505F and 260505J; Part No. 00055407504 UDI-DI 08426950807537 Lot 260505R; Part No. 00055415010 UDI-DI 08426950807698 Lots 260505H and 260505K; Part No. 00055430010 UDI-DI 08426950807759 Lots 260505H, 260505K and 260505F; Part No. 00055430011 UDI-DI 08426950807766 Lot 260505K; Product No. 00055445010 UDI-DI 08426950807810 Lots 260504AA and 260505K; Product No. 00055445011 UDI-DI 08426950807827 Lots 260505F and 260505J; Product No. 00055360010.A UDI-DI 08426950807599 Lot 260504AF. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Instrumentation Laboratory
Distributor
Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, KY, MN, MO, MT, OH, PA, VA, WI, WV and the countries of China, Germany, Italy.
Lot
Part No. 00055360010 UDI-DI 08426950807599 Lots 260505K, 260504AG, 260505F; Part No. 00055360011 UDI-DI 08426950807605 Lots 260505F and 260505J; Part No. 00055407504 UDI-DI 08426950807537 Lot 260505R; Part No. 00055415010 UDI-DI 08426950807698 Lots 260505H and 260505K; Part No. 00055430010 UDI-DI 08426950807759 Lots 260505H, 260505K and 260505F; Part No. 00055430011 UDI-DI 08426950807766 Lot 260505K; Product No. 00055445010 UDI-DI 08426950807810 Lots 260504AA and 260505K; Product No. 00055445011 UDI-DI 08426950807827 Lots 260505F and 260505J; Product No. 00055360010.A UDI-DI 08426950807599 Lot 260504AF;
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was GEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011 00055445010 00055445011 00055360010.A recalled?

Yes. FDA published recall Z-2900-2026 on August 12, 2026. Reported hazard: The active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.

What should I do if I have this product?

Voluntary: Firm initiated · Letter