Cobra x5 1-CH EMG ET Tube, 6.0mm, REF: LTE700S-5, Rx Only, Sterile EO UDI: B006:TE700S52$$7050918B - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.

Published September 2, 2020 · Recall initiated: July 21, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Neurovision Medical Products Inc
Distributor
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, MA, MD, MI, MN, MO, MS, NE, NY, OH, OR, PA, SC, SD, TN, TX, WA, WI and the countries of Canada, Dominican Republic, Hong Kong.
Lot
Kit Lots: 073119A 082019B 022820A 041620C 062520A
Risk classification
Class II
Area (FDA)
device
Country
United States