Cobra x5 1-Ch EMG ET Tube, 7.0mm, REF LTE700M-5, Rx Only, Sterile EO CE UDI: B006LTE700M52/$$7061418C - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.
Published September 2, 2020 · Recall initiated: July 21, 2020
Summary
FDA records medical device Class II recall Z-2904-2020, published September 2, 2020 and initiated July 21, 2020. Recalling firm: Neurovision Medical Products Inc; Status: Terminated; Reported quantity: N/A; Lot: Kit Lots: 052219E 052219D 052319A 052919C 053119C 060319A 060619A 060719C 061119A 061419A 061919A 062119A 070119C 071219B 080819B 082719E 091219A 091219B 091319A 091619C 091619D 012920A 020420D 020520B 020620A 020620D 020620G 021020E 021420A 021720D 022020C 022420E 030920C 030920D 031620A 031620B 041620B 051920C 060820E 071719A 071819D 072519A 072519C 080819C 081319A 081919A 082619E 091819B 091919A 091919F 100119A 101019A 101019C 101419D 101719A 110519A 110519B 110519C 110719F 012220A 012220E 012320B 012420D 012820A 012820C 042419A 042519B 042519A 050319A 050619A 050719A 051419C 052119B 082619D 112719G 112719H 120219B 120319B 120619C 051220A 051920A 052620A 052820C 060120C 061120A 061220B. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Neurovision Medical Products Inc
- Distributor
- Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, MA, MD, MI, MN, MO, MS, NE, NY, OH, OR, PA, SC, SD, TN, TX, WA, WI and the countries of Canada, Dominican Republic, Hong Kong.
- Lot
- Kit Lots: 052219E 052219D 052319A 052919C 053119C 060319A 060619A 060719C 061119A 061419A 061919A 062119A 070119C 071219B 080819B 082719E 091219A 091219B 091319A 091619C 091619D 012920A 020420D 020520B 020620A 020620D 020620G 021020E 021420A 021720D 022020C 022420E 030920C 030920D 031620A 031620B 041620B 051920C 060820E 071719A 071819D 072519A 072519C 080819C 081319A 081919A 082619E 091819B 091919A 091919F 100119A 101019A 101019C 101419D 101719A 110519A 110519B 110519C 110719F 012220A 012220E 012320B 012420D 012820A 012820C 042419A 042519B 042519A 050319A 050619A 050719A 051419C 052119B 082619D 112719G 112719H 120219B 120319B 120619C 051220A 051920A 052620A 052820C 060120C 061120A 061220B
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACobra, 3-Plate EMG ET Tube, 6.0mm, REF: LTE7003PM, not made with natural rubber latex, Rx Only, STERILEEOSeptember 4, 2024
- FDAJaloma Antiseptic Hand Sanitizer, Ethyl Alcohol 62% with Vitamin E, packaged in a) 16.9 Fl. Oz. (1.06 pint,September 9, 2020
- FDANerveana x5 RLN Monitoring Kit with Cobra 1-Ch EMG ET Tube, 8.0 mm, REF NVTKIT-C8-5, Rx Only, SterileSeptember 2, 2020
- FDACobra x5 2-Ch EMG ET Tube, 7.0mm, REF: LTE700DCM-5, Rx Only, Sterile EO UDI: B006LTE700DCM52 - Product Usage:September 2, 2020
- FDACobra x5 1-CH EMG ET Tube, 6.0mm, REF: LTE700S-5, Rx Only, Sterile EO UDI: B006:TE700S52$$7050918B - ProductSeptember 2, 2020
- FDAE.cam or Symbia systems that use foresight detectors - Product Usage: To detect or image the distribution ofSeptember 2, 2020
Questions about this recall
Was Cobra x5 1-Ch EMG ET Tube, 7.0mm, REF LTE700M-5, Rx Only, Sterile EO CE UDI: B006LTE700M52/$$7061418C - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system. recalled?
Yes. FDA published recall Z-2904-2020 on September 2, 2020. Reported hazard: The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.
What should I do if I have this product?
Voluntary: Firm initiated · Letter