Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO3010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
Published August 12, 2026 · Recall initiated: June 15, 2026
Summary
FDA records medical device Class II recall Z-2905-2026, published August 12, 2026 and initiated June 15, 2026. Recalling firm: TISSIUM SA; Status: Ongoing; Lot: Lot Code: Model No NCOCO3010; UDI-DI 03760279380117; Lot Numbers 250700045, 251000040, 251000036, 251000037, 251000039, 251000035, and 251000064. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- TISSIUM SA
- Distributor
- US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
- Lot
- Lot Code: Model No NCOCO3010; UDI-DI 03760279380117; Lot Numbers 250700045, 251000040, 251000036, 251000037, 251000039, 251000035, and 251000064
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- France
Related recalls
- FDABrand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO6010 Software Version:August 12, 2026
- FDABrand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2010 Software Version:August 12, 2026
- FDABrand Name: COAPTIUM Connect Product Name: COAPTIUM Connect Model/Catalog Number: NCOCO1510 Software Version:August 12, 2026
- FDABrand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO1575 Software Version:August 12, 2026
- FDABrand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version:August 12, 2026
- FDABrand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO4010 Software Version:August 12, 2026
Questions about this recall
Was Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO3010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No recalled?
Yes. FDA published recall Z-2905-2026 on August 12, 2026. Reported hazard: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
What should I do if I have this product?
Voluntary: Firm initiated · Letter