Back to search
FDAFDA · Food and drugs·Recall number Z-2901-2026

Brand Name: COAPTIUM Connect Product Name: COAPTIUM Connect Model/Catalog Number: NCOCO1510 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: N/A

Published August 12, 2026 · Recall initiated: June 15, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
TISSIUM SA
Distributor
US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
Lot
Lot Code: Model No NCOCO1510; UDI-DI 03760279380070; Lot Numbers 250700041 and 251000032
Risk classification
Class II
Area (FDA)
device
Country
France