Back to search
FDAFDA · Food and drugs·Recall number Z-2916-2026

Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A Product Description: The Spacelabs Healthcare (Spacelabs) Xhibit Central Station, Model 96102. Xhibit offers remote surveillance of patient data for those patients connected to a Spacelabs Healthcare bedside monitors and telemetry systems. The DisplayPort audio/video splitter allows users to extend the audio/video source from the Xhibit to additional external displays or monitors. This extension of audio/video is for viewing only and does not allow interaction with the Xhibit. Component: The DisplayPort audio/video splitter is a 3rd party external accessory to Xhibit.

Published August 19, 2026 · Recall initiated: June 4, 2026

Summary

FDA records medical device Class II recall Z-2916-2026, published August 19, 2026 and initiated June 4, 2026. Recalling firm: Spacelabs Healthcare, Inc.; Status: Ongoing; Reported quantity: 6074; Lot: Model/Catalog Number: 96102 DisplayPort Splitter Part Number: 010-1975-00 UDI-DI codes: 1084152212706 10841522100345 Serial Numbers: Pending. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to off-the-shelf display port component that has audio/video failures

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Spacelabs Healthcare, Inc.
Distributor
Worldwide - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WY and the countries of CANADA, FRANCE, ITALY, MALAYSIA, THAILAND.
Lot
Model/Catalog Number: 96102 DisplayPort Splitter Part Number: 010-1975-00 UDI-DI codes: 1084152212706 10841522100345 Serial Numbers: Pending
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A Product Description: The Spacelabs Healthcare (Spacelabs) Xhibit Central Station, Model 96102. Xhibit offers remote surveillance of patient data for those patients connected to a Spacelabs Healthcare bedside monitors and telemetry systems. The DisplayPort audio/video splitter allows users to extend the audio/video source from the Xhibit to additional external displays or monitors. This extension of audio/video is for viewing only and does not allow interaction with the Xhibit. Component: The DisplayPort audio/video splitter is a 3rd party external accessory to Xhibit. recalled?

Yes. FDA published recall Z-2916-2026 on August 19, 2026. Reported hazard: Due to off-the-shelf display port component that has audio/video failures

What should I do if I have this product?

Voluntary: Firm initiated · Letter