The TaqPath RT-PCR COVID-19 Kit with Applied Biosystem's COVID-19 Interpretive Software v2.0 or v2.2 (used with the QuantStudio Real-Time PCR platforms), includes the assays and controls for a multiplex real-time RT-PCR test for the qualitative detection of RNA from SARS-CoV-2 in upper respiratory specimens and bronchoalveolar lavage specimens from individuals suspected of COVID-19 by their healthcare provider. The test is interpreted by COVID-19 Interpretive Software

Published September 23, 2020 · Recall initiated: August 6, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

COVID-19 Interpretive Software issues that may cause: 1) the risk of invalid or inconclusive tests/false negative and potentially increasing the retesting burden on customers. 2) false positive results due to improper vortexing.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Life Technologies Corporation
Distributor
US: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NJ, NM, NY, NV, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. OUS: Afghanistan, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Botswana, Brazil, Cameroon, Canada, Cypress, Estonia, France, Georgia, Germany, Greece, Guadeloupe, Hungary, India, Indonesia, Ireland, Israel, Italy, Kenya, Kuwait, Lebanon, Mozambique, Namibia, New Zealand, Oman, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom, United Republic of Tanzania, and Vietnam.
Lot
Product SKU: Product Lot/Serial#: Release Date: 100093765 Version 1.2 (EUA) 28-APR-20 100087427 Version 2.2 (EUA) 11-MAY-20 100093771 Version 1.2 (CE-IVD) 28-APR-20 100094318 Version 2.2 (CE-IVD) 19-MAY-20 Applied Biosystems" COVID-19 Interpretive Software Version 1.2 was the first version of software developed in a dedicated CE-IVD Edition. Previous versions are shared with the FDA-EUA version of the device.
Risk classification
Class II
Area (FDA)
device
Country
United States