Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
Published August 26, 2026 · Recall initiated: July 8, 2026
Summary
FDA records medical device Class II recall Z-2983-2026, published August 26, 2026 and initiated July 8, 2026. Recalling firm: Fort Defiance Industries, LLC; Status: Ongoing; Lot: Lot Code: All serial numbers/ GTIN: (01) 00862992000327. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Fort Defiance Industries, LLC
- Distributor
- Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
- Lot
- Lot Code: All serial numbers/ GTIN: (01) 00862992000327
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLWAugust 26, 2026
- FDAEsteem Body Soft Convex Urostomy Pouch Model/Catalog Numbers: 423703, 423704, 423705, 423707, 423708, 423709,August 26, 2026
- FDAMiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;August 26, 2026
- FDABrand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid BaseplateAugust 26, 2026
- FDAMiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;August 26, 2026
- FDAAmerican Contract Systems medical procedure kits and trays labeled as: 1) CRANI PACK - 208400, Model ItemAugust 26, 2026
Questions about this recall
Was Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131 recalled?
Yes. FDA published recall Z-2983-2026 on August 26, 2026. Reported hazard: The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
What should I do if I have this product?
Voluntary: Firm initiated