Back to search
FDAFDA · Food and drugs·Recall number Z-2983-2026

Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

Published August 26, 2026 · Recall initiated: July 8, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Fort Defiance Industries, LLC
Distributor
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
Lot
Lot Code: All serial numbers/ GTIN: (01) 00862992000327
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card