FDAFDA · Food and drugs·Recall number Z-2983-2026
Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
Published August 26, 2026 · Recall initiated: July 8, 2026
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
What to do
Voluntary: Firm initiated
Recall details
- Manufacturer
- Fort Defiance Industries, LLC
- Distributor
- Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
- Lot
- Lot Code: All serial numbers/ GTIN: (01) 00862992000327
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States