FDAFDA · Food and drugs·Recall number Z-2983-2026

Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

Published August 26, 2026 · Recall initiated: July 8, 2026

Summary

FDA records medical device Class II recall Z-2983-2026, published August 26, 2026 and initiated July 8, 2026. Recalling firm: Fort Defiance Industries, LLC; Status: Ongoing; Lot: Lot Code: All serial numbers/ GTIN: (01) 00862992000327. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.

What to do

Voluntary: Firm initiated

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Fort Defiance Industries, LLC
Distributor
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
Lot
Lot Code: All serial numbers/ GTIN: (01) 00862992000327
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131 recalled?

Yes. FDA published recall Z-2983-2026 on August 26, 2026. Reported hazard: The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.

What should I do if I have this product?

Voluntary: Firm initiated