Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate Model/Catalog Number: 1500-1030; 1500-1035 Software Version: N/A - off-the-shelf hardware product Product Description: The restor3d Reverse Total Shoulder Arthroplasty System is intended for patients requiring a reverse shoulder replacement for patients with a functional deltoid muscle and with a deficient rotator cuff. The restor3d Reverse Total Shoulder Arthroplasty System consists of the glenoid baseplate, glenosphere intraoperatively affixed to the baseplate, humeral stem, and polymer bearing component affixed to the humeral stem. Additionally, the system includes supporting standard instrument trays and all required accessories (e.g., security screws, peripheral screws). The restor3d rTSA System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The Veritas standard glenoid baseplate is the only component within scope of this field action. The patient-specific baseplate is unaffected by this defect and therefore not in scope. Component: Reverse Total Shoulder
Published August 26, 2026 · Recall initiated: June 30, 2026
Summary
FDA records medical device Class II recall Z-2987-2026, published August 26, 2026 and initiated June 30, 2026. Recalling firm: restor3d Inc.; Status: Ongoing; Reported quantity: 59; Lot: Lot Code: Model No 1500-1030; UDI-DI 00840097513225; Lot Numbers A00030, A00065, A09828, A09830 Model No 1500-1035; UDI-DI 00840097513232; Lot Numbers A00064, A09829. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to incorrect machining of peripheral screw hole resulting in potential for improper engagement of the peripheral screws.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- restor3d Inc.
- Distributor
- U.S. Nationwide distribution in the state of Texas.
- Lot
- Lot Code: Model No 1500-1030; UDI-DI 00840097513225; Lot Numbers A00030, A00065, A09828, A09830 Model No 1500-1035; UDI-DI 00840097513232; Lot Numbers A00064, A09829
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVeritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1,November 11, 2026
- FDAGeneric Name: restor3d MTP Implant Generic Name: MTP Implant Model/Catalog Number: 1103-161620 SoftwareOctober 28, 2026
- FDAKinos Axiom Total Ankle System with Axiom PSR, Flat-cut (Cut Guide), REF: 5831-9006September 23, 2026
- FDAEsteem Body Soft Convex Urostomy Pouch Model/Catalog Numbers: 423703, 423704, 423705, 423707, 423708, 423709,August 26, 2026
- FDAAmerican Contract Systems medical procedure kits and trays labeled as: 1) CRANI PACK - 208400, Model ItemAugust 26, 2026
- FDAFort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131August 26, 2026
Questions about this recall
Was Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate Model/Catalog Number: 1500-1030; 1500-1035 Software Version: N/A - off-the-shelf hardware product Product Description: The restor3d Reverse Total Shoulder Arthroplasty System is intended for patients requiring a reverse shoulder replacement for patients with a functional deltoid muscle and with a deficient rotator cuff. The restor3d Reverse Total Shoulder Arthroplasty System consists of the glenoid baseplate, glenosphere intraoperatively affixed to the baseplate, humeral stem, and polymer bearing component affixed to the humeral stem. Additionally, the system includes supporting standard instrument trays and all required accessories (e.g., security screws, peripheral screws). The restor3d rTSA System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The Veritas standard glenoid baseplate is the only component within scope of this field action. The patient-specific baseplate is unaffected by this defect and therefore not in scope. Component: Reverse Total Shoulder recalled?
Yes. FDA published recall Z-2987-2026 on August 26, 2026. Reported hazard: Due to incorrect machining of peripheral screw hole resulting in potential for improper engagement of the peripheral screws.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail