Back to search
FDAFDA · Food and drugs·Recall number Z-3044-2020

TDDA Drugs of Abuse Cup(AMT Item/ AMT Description): TDDA-6MBAU-CUP /"TDDA Drugs of Abuse Cup AMP300/BZO200/COC300/MET300/OPI300/ THC50 with Cr, Ni, pH, Bl, S.G."

Published October 7, 2020 · Recall initiated: August 18, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Ameditech Inc
Distributor
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
Lot
ALL Lots Within Expiry
Risk classification
Class II
Area (FDA)
device
Country
United States