FDAFDA · Food and drugs·Recall number Z-3053-2020
AC/DC POWER SUPPLY for 7400, Part Number 151004301, XP Power Model Number AHM180PS19-XE1057, AC/DC Adapter associated with the following portable ultrasound diagnostic systems: The AC/DC adapter is intended for use with the following portable ultrasound systems: MyLabAlpha (101740000) MyLabGamma (101741000) MyLabOmega (101740004) MyLabSigma (101741004) MyLabSigma elite (101741007) MyLabAlphaVET (101740001) MyLabGammaVET (101741001) MyLabDelta (100743000) MyLabDeltaVET (100743001) MyLabOmegaVET (101740005) MyLabSigmaVET (101741005)
Published October 7, 2020 · Recall initiated: September 8, 2020
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting. This condition is caused by the mechanical stress of the connection between the AC/DC power adapter and the mains cable over the time of use.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- ESAOTE S.P.A
- Distributor
- US Nationwide
- Lot
- Serial Numbers: Below K17350001
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Italy