LX3 Floor Stand for LuxOR Ophthalmic Microscope; Endure LuxOR Q-VUE Ophthalmic Microscope. LX3 Microscope Floor Stands are not stand-alone medical devices, but are used in combination with the LuxOR Ophthalmic Microscope and the LuxOR Ophthalmic Microscope with Q-VUE 3-D Assistant Visualization. These microscopes are used for low magnification visualization during cataract, retina, and cornea surgeries

Published September 6, 2017 · Recall initiated: May 25, 2017

Summary

FDA records medical device Class II recall Z-3081-2017, published September 6, 2017 and initiated May 25, 2017. Recalling firm: Alcon Research, Ltd.; Status: Terminated; Reported quantity: 1,074 units; Lot: Domestic Serial Numbers: 1402548501X, 1402548502X, 1402548503X, 1402548504X, 1402548505X, 1402548506X, 1402577401X, 1402577402X, 1402577403X, 1402577404X, 1402577405X, 1402577406X, 1402612001X, 1402612002X, 1402612003X, 1402612004X, 1402612005X, 1402612006X, 1402655703X, 1402655704X, 1402655705X, 1402655706X, 1402705203X, 1402705204X, 1402728703X, 1402728704X, 1402728705X, 1402728706X, 1402804805X, 1402804806X, 1402887601X, 1402887602X, 1402887603X, 1402887604X, 1402959002X, 1402959004X, 1402959005X, 1402959006X, 1403012601X, 1403012602X, 1403012603X, 1403012604X, 1403012605X, 1403012606X, 1403045903X, 1403045904X, 1403045905X, 1403045906X, 1403072801X, 1403072802X, 1403072803X, 1403072804X, 1403072805X, 1403072806X, 1403124901X, 1403124902X, 1403124903X, 1403124904X, 1403124905X, 1403124906X, 1403145301X, 1403145302X, 1403145303X, 1403145304X, 1403145305X, 1403145306X, 1403173401X, 1403173402X, 1403173403X, 1403173404X, 1403173405X, 1403173406X, 1403211701X, 1403211702X, 1403211704X, 1403211706X, 1403239601X, 1403239602X, 1403239603X, 1403239604X, 1403239605X, 1403239606X, 1403253702X, 1403253703X, 1403253705X, 1403253706X, 1403272701X, 1403272702X, 1403351301X, 1403351302X, 1403351303X, 1403351304X, 1403351305X, 1403351306X, 1403390506X, 1403390803X, 1403390805X, 1403440402X, 1403440406X, 1403440502X, 1403440503X, 1403440505X, 1403440506X, 1403484502X, 1403484602X, 1403484603X, 1403484604X, 1403539702X, 1403539703X, 1403539704X, 1403539706X, 1403625401X, 1403625402X, 1403625403X, 1403625404X, 1403625405X, 1403625406X, 1403625501X, 1403625502X, 1403625504X, 1403625505X, 1403688701X, 1403688702X, 1403688801X, 1403688802X, 1403688803X, 1403688804X, 1403688805X, 1403688806X, 1403736301X, 1403736302X, 1403736303X, 1403736304X, 1403736305X, 1403736306X, 1403736401X, 1403736404X, 1403791801X, 1403791803X, 1403791805X, 1403791806X, 1403830904X, 1403830905X, 1403830906X, 1403831004X, 1403831005X, 1403831006X, 1403862204X, 1403862205X, 1403862206X, 1403862305X, 1403862306X, 1403886401X, 1403886402X, 1403886403X, 1403886404X, 1403886405X, 1403886406X, 1403886501X, 1403886502X, 1403886503X, 1403886504X, 1403886505X, 1403886506X, 1403986603X, 1403986604X, 1403986605X, 1403986606X, 1403990601X, 1403990602X, 1403990603X, 1403990604X, 1403990701X, 1403990703X, 1403990704X, 1403990705X, 1403990706X, 1404022104X, 1404022105X, 1404022202X, 1404022204X, 1501030401X, 1501030601X, 1501030606X, 1501088001X, 1501088006X, 1501088101X, 1501088105X, 1501088106X, 1501132903X, 1501132905X, 1501132906X, 1501158301X, 1501158302X, 1501158303X, 1501158304X, 1501158401X, 1501158402X, 1501158403X, 1501158404X, 1501158406X, 1501201501X, 1501201506X, 1501274503X, 1501355602X, 1501355603X, 1501355604X, 1501355605X, 1501355606X, 1501355702X, 1501565905X, 1501566001X, 1501566003X, 1501566004X, 150156600SX, 1501566006X, 1501619502X, 1501619503X, 1501619506X, 1501619601X, 1501619604X, 150161960SX, 1501619606X, 1501646804X, 1501646902X, 1501646903X, 1501646906X, 1501672006X, 1501672102X, 1501672105X, 1501672106X, 1501731601X, 1501731602X, 1501731604X, 1501731805X, 1501731806X, 1501790001X, 1501790002X, 1501790003X, 1501790004X, 1501790005X, 1501790006X, 1501790103X, 1501790104X, 1501790105X, 1501790106X, 1501838401X, 1501838402X, 1501838403X, 1501838404X, 1501838405X, 1501838406X, 1501898901X, 1501898902X, 1501898903X, 1501898904X, 1501898905X, 1501899003X, 1501899006X, 1501940101X, 1501940102X, 1501940104X, 1501940105X, 1501940106X, 1501955601X, 1501955602X, 1501955603X, 1501955604X, 1501955605X, 1502051002X, 1502051004X, 1502051006X, 1502134606X, 1502134701X, 1502134703X, 1502134705X, 1502176206X, 1502326701X, 1502326702X, 1502384602X, 1502384603X, 1502384605X, 1502384606X, 1502384705X, 1502384706X, 1502443501X, 1502443503X, 1502443504X, 1502443506X, 1502459501X, 1502459502X, 1502459601X, 1502459603X, 1502459604X, 1502473801X, 1502473802X, 1502473803X, 1502473804X, 1502473805X, 1502473806X, 1502473904X, 1502473905X, 1502499501X, 1502499504X, 1502499505X, . Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Multiple shipments or forceful impact to a caster on a LuxOr/LX3 microscope may stress the microscope base, which could contribute to a caster breaking or detaching from the base.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Alcon Research, Ltd.
Distributor
Worldwide Distribution - US (nationwide and Puerto Rico): Units were also distributed globally to the following countries: Algeria, Argentina, Australia, Austria, Belarus, Belgium, Bosnia, Herzegovina, Bulgaria, Canada, Chile, Columbia, Costa Rica, Croatia, Curacao, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Ecuador, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hungary, India, Italy, Jamaica, Japan, Jordan, Kazakhstan, South Korea, Kuwait, Lebanon, Lithuania, Luxembourg, Malaysia, Mexico, Mongolia, Morocco, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, Ukraine, United Kingdom, and Uzbekistan
Lot
Domestic Serial Numbers: 1402548501X, 1402548502X, 1402548503X, 1402548504X, 1402548505X, 1402548506X, 1402577401X, 1402577402X, 1402577403X, 1402577404X, 1402577405X, 1402577406X, 1402612001X, 1402612002X, 1402612003X, 1402612004X, 1402612005X, 1402612006X, 1402655703X, 1402655704X, 1402655705X, 1402655706X, 1402705203X, 1402705204X, 1402728703X, 1402728704X, 1402728705X, 1402728706X, 1402804805X, 1402804806X, 1402887601X, 1402887602X, 1402887603X, 1402887604X, 1402959002X, 1402959004X, 1402959005X, 1402959006X, 1403012601X, 1403012602X, 1403012603X, 1403012604X, 1403012605X, 1403012606X, 1403045903X, 1403045904X, 1403045905X, 1403045906X, 1403072801X, 1403072802X, 1403072803X, 1403072804X, 1403072805X, 1403072806X, 1403124901X, 1403124902X, 1403124903X, 1403124904X, 1403124905X, 1403124906X, 1403145301X, 1403145302X, 1403145303X, 1403145304X, 1403145305X, 1403145306X, 1403173401X, 1403173402X, 1403173403X, 1403173404X, 1403173405X, 1403173406X, 1403211701X, 1403211702X, 1403211704X, 1403211706X, 1403239601X, 1403239602X, 1403239603X, 1403239604X, 1403239605X, 1403239606X, 1403253702X, 1403253703X, 1403253705X, 1403253706X, 1403272701X, 1403272702X, 1403351301X, 1403351302X, 1403351303X, 1403351304X, 1403351305X, 1403351306X, 1403390506X, 1403390803X, 1403390805X, 1403440402X, 1403440406X, 1403440502X, 1403440503X, 1403440505X, 1403440506X, 1403484502X, 1403484602X, 1403484603X, 1403484604X, 1403539702X, 1403539703X, 1403539704X, 1403539706X, 1403625401X, 1403625402X, 1403625403X, 1403625404X, 1403625405X, 1403625406X, 1403625501X, 1403625502X, 1403625504X, 1403625505X, 1403688701X, 1403688702X, 1403688801X, 1403688802X, 1403688803X, 1403688804X, 1403688805X, 1403688806X, 1403736301X, 1403736302X, 1403736303X, 1403736304X, 1403736305X, 1403736306X, 1403736401X, 1403736404X, 1403791801X, 1403791803X, 1403791805X, 1403791806X, 1403830904X, 1403830905X, 1403830906X, 1403831004X, 1403831005X, 1403831006X, 1403862204X, 1403862205X, 1403862206X, 1403862305X, 1403862306X, 1403886401X, 1403886402X, 1403886403X, 1403886404X, 1403886405X, 1403886406X, 1403886501X, 1403886502X, 1403886503X, 1403886504X, 1403886505X, 1403886506X, 1403986603X, 1403986604X, 1403986605X, 1403986606X, 1403990601X, 1403990602X, 1403990603X, 1403990604X, 1403990701X, 1403990703X, 1403990704X, 1403990705X, 1403990706X, 1404022104X, 1404022105X, 1404022202X, 1404022204X, 1501030401X, 1501030601X, 1501030606X, 1501088001X, 1501088006X, 1501088101X, 1501088105X, 1501088106X, 1501132903X, 1501132905X, 1501132906X, 1501158301X, 1501158302X, 1501158303X, 1501158304X, 1501158401X, 1501158402X, 1501158403X, 1501158404X, 1501158406X, 1501201501X, 1501201506X, 1501274503X, 1501355602X, 1501355603X, 1501355604X, 1501355605X, 1501355606X, 1501355702X, 1501565905X, 1501566001X, 1501566003X, 1501566004X, 150156600SX, 1501566006X, 1501619502X, 1501619503X, 1501619506X, 1501619601X, 1501619604X, 150161960SX, 1501619606X, 1501646804X, 1501646902X, 1501646903X, 1501646906X, 1501672006X, 1501672102X, 1501672105X, 1501672106X, 1501731601X, 1501731602X, 1501731604X, 1501731805X, 1501731806X, 1501790001X, 1501790002X, 1501790003X, 1501790004X, 1501790005X, 1501790006X, 1501790103X, 1501790104X, 1501790105X, 1501790106X, 1501838401X, 1501838402X, 1501838403X, 1501838404X, 1501838405X, 1501838406X, 1501898901X, 1501898902X, 1501898903X, 1501898904X, 1501898905X, 1501899003X, 1501899006X, 1501940101X, 1501940102X, 1501940104X, 1501940105X, 1501940106X, 1501955601X, 1501955602X, 1501955603X, 1501955604X, 1501955605X, 1502051002X, 1502051004X, 1502051006X, 1502134606X, 1502134701X, 1502134703X, 1502134705X, 1502176206X, 1502326701X, 1502326702X, 1502384602X, 1502384603X, 1502384605X, 1502384606X, 1502384705X, 1502384706X, 1502443501X, 1502443503X, 1502443504X, 1502443506X, 1502459501X, 1502459502X, 1502459601X, 1502459603X, 1502459604X, 1502473801X, 1502473802X, 1502473803X, 1502473804X, 1502473805X, 1502473806X, 1502473904X, 1502473905X, 1502499501X, 1502499504X, 1502499505X,
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was LX3 Floor Stand for LuxOR Ophthalmic Microscope; Endure LuxOR Q-VUE Ophthalmic Microscope. LX3 Microscope Floor Stands are not stand-alone medical devices, but are used in combination with the LuxOR Ophthalmic Microscope and the LuxOR Ophthalmic Microscope with Q-VUE 3-D Assistant Visualization. These microscopes are used for low magnification visualization during cataract, retina, and cornea surgeries recalled?

Yes. FDA published recall Z-3081-2017 on September 6, 2017. Reported hazard: Multiple shipments or forceful impact to a caster on a LuxOr/LX3 microscope may stress the microscope base, which could contribute to a caster breaking or detaching from the base.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit