LX3 Floor Stand for LuxOR Ophthalmic Microscope; Endure LuxOR Q-VUE Ophthalmic Microscope. LX3 Microscope Floor Stands are not stand-alone medical devices, but are used in combination with the LuxOR Ophthalmic Microscope and the LuxOR Ophthalmic Microscope with Q-VUE 3-D Assistant Visualization. These microscopes are used for low magnification visualization during cataract, retina, and cornea surgeries
Published September 6, 2017 · Recall initiated: May 25, 2017
Summary
FDA records medical device Class II recall Z-3081-2017, published September 6, 2017 and initiated May 25, 2017. Recalling firm: Alcon Research, Ltd.; Status: Terminated; Reported quantity: 1,074 units; Lot: Domestic Serial Numbers: 1402548501X, 1402548502X, 1402548503X, 1402548504X, 1402548505X, 1402548506X, 1402577401X, 1402577402X, 1402577403X, 1402577404X, 1402577405X, 1402577406X, 1402612001X, 1402612002X, 1402612003X, 1402612004X, 1402612005X, 1402612006X, 1402655703X, 1402655704X, 1402655705X, 1402655706X, 1402705203X, 1402705204X, 1402728703X, 1402728704X, 1402728705X, 1402728706X, 1402804805X, 1402804806X, 1402887601X, 1402887602X, 1402887603X, 1402887604X, 1402959002X, 1402959004X, 1402959005X, 1402959006X, 1403012601X, 1403012602X, 1403012603X, 1403012604X, 1403012605X, 1403012606X, 1403045903X, 1403045904X, 1403045905X, 1403045906X, 1403072801X, 1403072802X, 1403072803X, 1403072804X, 1403072805X, 1403072806X, 1403124901X, 1403124902X, 1403124903X, 1403124904X, 1403124905X, 1403124906X, 1403145301X, 1403145302X, 1403145303X, 1403145304X, 1403145305X, 1403145306X, 1403173401X, 1403173402X, 1403173403X, 1403173404X, 1403173405X, 1403173406X, 1403211701X, 1403211702X, 1403211704X, 1403211706X, 1403239601X, 1403239602X, 1403239603X, 1403239604X, 1403239605X, 1403239606X, 1403253702X, 1403253703X, 1403253705X, 1403253706X, 1403272701X, 1403272702X, 1403351301X, 1403351302X, 1403351303X, 1403351304X, 1403351305X, 1403351306X, 1403390506X, 1403390803X, 1403390805X, 1403440402X, 1403440406X, 1403440502X, 1403440503X, 1403440505X, 1403440506X, 1403484502X, 1403484602X, 1403484603X, 1403484604X, 1403539702X, 1403539703X, 1403539704X, 1403539706X, 1403625401X, 1403625402X, 1403625403X, 1403625404X, 1403625405X, 1403625406X, 1403625501X, 1403625502X, 1403625504X, 1403625505X, 1403688701X, 1403688702X, 1403688801X, 1403688802X, 1403688803X, 1403688804X, 1403688805X, 1403688806X, 1403736301X, 1403736302X, 1403736303X, 1403736304X, 1403736305X, 1403736306X, 1403736401X, 1403736404X, 1403791801X, 1403791803X, 1403791805X, 1403791806X, 1403830904X, 1403830905X, 1403830906X, 1403831004X, 1403831005X, 1403831006X, 1403862204X, 1403862205X, 1403862206X, 1403862305X, 1403862306X, 1403886401X, 1403886402X, 1403886403X, 1403886404X, 1403886405X, 1403886406X, 1403886501X, 1403886502X, 1403886503X, 1403886504X, 1403886505X, 1403886506X, 1403986603X, 1403986604X, 1403986605X, 1403986606X, 1403990601X, 1403990602X, 1403990603X, 1403990604X, 1403990701X, 1403990703X, 1403990704X, 1403990705X, 1403990706X, 1404022104X, 1404022105X, 1404022202X, 1404022204X, 1501030401X, 1501030601X, 1501030606X, 1501088001X, 1501088006X, 1501088101X, 1501088105X, 1501088106X, 1501132903X, 1501132905X, 1501132906X, 1501158301X, 1501158302X, 1501158303X, 1501158304X, 1501158401X, 1501158402X, 1501158403X, 1501158404X, 1501158406X, 1501201501X, 1501201506X, 1501274503X, 1501355602X, 1501355603X, 1501355604X, 1501355605X, 1501355606X, 1501355702X, 1501565905X, 1501566001X, 1501566003X, 1501566004X, 150156600SX, 1501566006X, 1501619502X, 1501619503X, 1501619506X, 1501619601X, 1501619604X, 150161960SX, 1501619606X, 1501646804X, 1501646902X, 1501646903X, 1501646906X, 1501672006X, 1501672102X, 1501672105X, 1501672106X, 1501731601X, 1501731602X, 1501731604X, 1501731805X, 1501731806X, 1501790001X, 1501790002X, 1501790003X, 1501790004X, 1501790005X, 1501790006X, 1501790103X, 1501790104X, 1501790105X, 1501790106X, 1501838401X, 1501838402X, 1501838403X, 1501838404X, 1501838405X, 1501838406X, 1501898901X, 1501898902X, 1501898903X, 1501898904X, 1501898905X, 1501899003X, 1501899006X, 1501940101X, 1501940102X, 1501940104X, 1501940105X, 1501940106X, 1501955601X, 1501955602X, 1501955603X, 1501955604X, 1501955605X, 1502051002X, 1502051004X, 1502051006X, 1502134606X, 1502134701X, 1502134703X, 1502134705X, 1502176206X, 1502326701X, 1502326702X, 1502384602X, 1502384603X, 1502384605X, 1502384606X, 1502384705X, 1502384706X, 1502443501X, 1502443503X, 1502443504X, 1502443506X, 1502459501X, 1502459502X, 1502459601X, 1502459603X, 1502459604X, 1502473801X, 1502473802X, 1502473803X, 1502473804X, 1502473805X, 1502473806X, 1502473904X, 1502473905X, 1502499501X, 1502499504X, 1502499505X, . Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Multiple shipments or forceful impact to a caster on a LuxOr/LX3 microscope may stress the microscope base, which could contribute to a caster breaking or detaching from the base.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Alcon Research, Ltd.
- Distributor
- Worldwide Distribution - US (nationwide and Puerto Rico): Units were also distributed globally to the following countries: Algeria, Argentina, Australia, Austria, Belarus, Belgium, Bosnia, Herzegovina, Bulgaria, Canada, Chile, Columbia, Costa Rica, Croatia, Curacao, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Ecuador, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hungary, India, Italy, Jamaica, Japan, Jordan, Kazakhstan, South Korea, Kuwait, Lebanon, Lithuania, Luxembourg, Malaysia, Mexico, Mongolia, Morocco, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, Ukraine, United Kingdom, and Uzbekistan
- Lot
- Domestic Serial Numbers: 1402548501X, 1402548502X, 1402548503X, 1402548504X, 1402548505X, 1402548506X, 1402577401X, 1402577402X, 1402577403X, 1402577404X, 1402577405X, 1402577406X, 1402612001X, 1402612002X, 1402612003X, 1402612004X, 1402612005X, 1402612006X, 1402655703X, 1402655704X, 1402655705X, 1402655706X, 1402705203X, 1402705204X, 1402728703X, 1402728704X, 1402728705X, 1402728706X, 1402804805X, 1402804806X, 1402887601X, 1402887602X, 1402887603X, 1402887604X, 1402959002X, 1402959004X, 1402959005X, 1402959006X, 1403012601X, 1403012602X, 1403012603X, 1403012604X, 1403012605X, 1403012606X, 1403045903X, 1403045904X, 1403045905X, 1403045906X, 1403072801X, 1403072802X, 1403072803X, 1403072804X, 1403072805X, 1403072806X, 1403124901X, 1403124902X, 1403124903X, 1403124904X, 1403124905X, 1403124906X, 1403145301X, 1403145302X, 1403145303X, 1403145304X, 1403145305X, 1403145306X, 1403173401X, 1403173402X, 1403173403X, 1403173404X, 1403173405X, 1403173406X, 1403211701X, 1403211702X, 1403211704X, 1403211706X, 1403239601X, 1403239602X, 1403239603X, 1403239604X, 1403239605X, 1403239606X, 1403253702X, 1403253703X, 1403253705X, 1403253706X, 1403272701X, 1403272702X, 1403351301X, 1403351302X, 1403351303X, 1403351304X, 1403351305X, 1403351306X, 1403390506X, 1403390803X, 1403390805X, 1403440402X, 1403440406X, 1403440502X, 1403440503X, 1403440505X, 1403440506X, 1403484502X, 1403484602X, 1403484603X, 1403484604X, 1403539702X, 1403539703X, 1403539704X, 1403539706X, 1403625401X, 1403625402X, 1403625403X, 1403625404X, 1403625405X, 1403625406X, 1403625501X, 1403625502X, 1403625504X, 1403625505X, 1403688701X, 1403688702X, 1403688801X, 1403688802X, 1403688803X, 1403688804X, 1403688805X, 1403688806X, 1403736301X, 1403736302X, 1403736303X, 1403736304X, 1403736305X, 1403736306X, 1403736401X, 1403736404X, 1403791801X, 1403791803X, 1403791805X, 1403791806X, 1403830904X, 1403830905X, 1403830906X, 1403831004X, 1403831005X, 1403831006X, 1403862204X, 1403862205X, 1403862206X, 1403862305X, 1403862306X, 1403886401X, 1403886402X, 1403886403X, 1403886404X, 1403886405X, 1403886406X, 1403886501X, 1403886502X, 1403886503X, 1403886504X, 1403886505X, 1403886506X, 1403986603X, 1403986604X, 1403986605X, 1403986606X, 1403990601X, 1403990602X, 1403990603X, 1403990604X, 1403990701X, 1403990703X, 1403990704X, 1403990705X, 1403990706X, 1404022104X, 1404022105X, 1404022202X, 1404022204X, 1501030401X, 1501030601X, 1501030606X, 1501088001X, 1501088006X, 1501088101X, 1501088105X, 1501088106X, 1501132903X, 1501132905X, 1501132906X, 1501158301X, 1501158302X, 1501158303X, 1501158304X, 1501158401X, 1501158402X, 1501158403X, 1501158404X, 1501158406X, 1501201501X, 1501201506X, 1501274503X, 1501355602X, 1501355603X, 1501355604X, 1501355605X, 1501355606X, 1501355702X, 1501565905X, 1501566001X, 1501566003X, 1501566004X, 150156600SX, 1501566006X, 1501619502X, 1501619503X, 1501619506X, 1501619601X, 1501619604X, 150161960SX, 1501619606X, 1501646804X, 1501646902X, 1501646903X, 1501646906X, 1501672006X, 1501672102X, 1501672105X, 1501672106X, 1501731601X, 1501731602X, 1501731604X, 1501731805X, 1501731806X, 1501790001X, 1501790002X, 1501790003X, 1501790004X, 1501790005X, 1501790006X, 1501790103X, 1501790104X, 1501790105X, 1501790106X, 1501838401X, 1501838402X, 1501838403X, 1501838404X, 1501838405X, 1501838406X, 1501898901X, 1501898902X, 1501898903X, 1501898904X, 1501898905X, 1501899003X, 1501899006X, 1501940101X, 1501940102X, 1501940104X, 1501940105X, 1501940106X, 1501955601X, 1501955602X, 1501955603X, 1501955604X, 1501955605X, 1502051002X, 1502051004X, 1502051006X, 1502134606X, 1502134701X, 1502134703X, 1502134705X, 1502176206X, 1502326701X, 1502326702X, 1502384602X, 1502384603X, 1502384605X, 1502384606X, 1502384705X, 1502384706X, 1502443501X, 1502443503X, 1502443504X, 1502443506X, 1502459501X, 1502459502X, 1502459601X, 1502459603X, 1502459604X, 1502473801X, 1502473802X, 1502473803X, 1502473804X, 1502473805X, 1502473806X, 1502473904X, 1502473905X, 1502499501X, 1502499504X, 1502499505X,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAirLife(TM) Reduced Heparin ARTERIAL BLOOD SAMPLER, REF 9025RHTR, 1ML SYRINE 25G 5/8", LUER SLIP PROTECTIVESeptember 6, 2017
- FDAArtis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family ofSeptember 6, 2017
- FDAORA System with VerifEye+ Cart, Catalog Number 8065998307 For use during intraocular lens surgeryAugust 30, 2017
- FDAORA System with VerifEye, Catalog Number 8065998300 For use during intraocular lens surgeryAugust 30, 2017
- FDA23G and 25G Constellation Totalplus Paks provides access into the posterior segment of the eye for surgicalMay 25, 2016
- FDACustom Paks containing 23G and 25G valved trocar cannula provide access into the posterior segment of the eyeMay 25, 2016
Questions about this recall
Was LX3 Floor Stand for LuxOR Ophthalmic Microscope; Endure LuxOR Q-VUE Ophthalmic Microscope. LX3 Microscope Floor Stands are not stand-alone medical devices, but are used in combination with the LuxOR Ophthalmic Microscope and the LuxOR Ophthalmic Microscope with Q-VUE 3-D Assistant Visualization. These microscopes are used for low magnification visualization during cataract, retina, and cornea surgeries recalled?
Yes. FDA published recall Z-3081-2017 on September 6, 2017. Reported hazard: Multiple shipments or forceful impact to a caster on a LuxOr/LX3 microscope may stress the microscope base, which could contribute to a caster breaking or detaching from the base.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit