ORA System with VerifEye+ Cart, Catalog Number 8065998307 For use during intraocular lens surgery

Published August 30, 2017 · Recall initiated: June 30, 2017

Summary

FDA records medical device Class II recall Z-3049-2017, published August 30, 2017 and initiated June 30, 2017. Recalling firm: Alcon Research, Ltd.; Status: Terminated; Reported quantity: 429 units; Lot: Serial numbers: C5013,C5014,C5015,C5016,C5017,C5018,C5019,C5020,C5021,C5023,C5024,C5025,C5026,C5027,C5028,C5029,C5030,C5031,C5032,C5033,C5034,C5035,C5036,C5038,C5039,C5040,C5041,C5042,C5043,C5044, C5045,C5046,C5047,C5048,C5049,C5050,C5051,C5053,C5054,C5055,C5059,C5060,C5061,C5062,C5064, C5065,C5066,C5069,C5071,C5075,C5076,C5077,C5078,C5079,C5080,C5081,C5082,C5083,C5090,C5093, C5094,C5097,C5098,C5101,C5102,C5105,C5106,C5107,C5108,C5109,C5110,C5111,C5112,C5113,C5116, C5117,C5119,C5120,C5121,C5122,C5125,C5126,C5127,C5128,C5129,C5130,C5131,C5132,C5133,C5134, C5136,C5137,C5138,C5139,C5140,C5141,C5142,C5143,C5144,C5146,C5147,C5148,C5149,C5150,C5151, C5152,C5153,C5154,C5155,C5156,C5157,C5158,C5159,C5160,C5161,C5163,C5164,C5165,C5166,C5167, C5168,C5169,C5170,C5171,C5173,C5174,C5175,C5176,C5177,C5178,C5179,C5180,C5181,C5182,C5183, C5184,C5185,C5186,C5187,C5188,C5189,C5190,C5191,C5192,C5193,C5194,C5195,C5196,C5197,C5198, C5199,C5200,C5201,C5202,C5203,C5204,C5205,C5207,C5214,C5215,C5216,C5217,C5218,C5219,C5220, C5221,C5222,C5223,C5224,C5225,C5226,C5227,C5228,C5229,C5230,C5231,C5232,C5233,C5238,C5239, C5240,C5241,C5242,C5243,C5244,C5245,C5246,C5248,C5249,C5250,C5251,C5252,C5253,C5255,C5256, C5257,C5258,C5259,C5260,C5261,C5262,C5263,C5264,C5265,C5266,C5267,C5268,C5269,C5270,C5271, C5272,C5273,C5274,C5276,C5277,C5278,C5279,C5280,C5281,C5282,C5283,C5284,C5285,C5286,C5287, C5288,C5289,C5290,C5291,C5292,C5293,C5294,C5295,C5296,C5297,C5298,C5299,C5300,C5301,C5302, C5303,C5304,C5305,C5306,C5307,C5308,C5309,C5310,C5311,C5312,C5313,C5314,C5315,C5316,C5317, C5318,C5319,C5320,C5321,C5322,C5323,C5326,C5328,C5329,C5330,C5334,C5335,C5336,C5339,C5340, C5341,C5343,C5348,C5349,C5350,C5352,C5353,C5354,C5355,C5356,C5357,C5358,C5359,C5360,C5361, C5362,C5363,C5364,C5365,C5366,C5368,C5369,C5370,C5371,C5372,C5373,C5374,C5375,C5376,C5377, C5378,C5379,C5380,C5381,C5382,C5383,C5384,C5385,C5386,C5387,C5388,C5389,C5390,C5391,C5392, C5393,C5394,C5395,C5396,C5397,C5398,C5399,C5400,C5401,C5402,C5403,C5404,C5405,C5406,C5407, C5408,C5409,C5410,C5411,C5412,C5413,C5414,C5415,C5416,C5417,C5418,C5434,C5435,C5436,C5437, C5438,C5439,C5440,C5441,C5442,C5443,C5444,C5446,C5448,C5449,C5451,C5452,C5453,C5454,C5456, C5458,C5462,C5463,C5464,C5465,C5466,C5467,C5469,C5471,C5473,C5475,C5476,C5477,C5478,C5479, C5484,C5485,C5486,C5487,C5488,C5489,C5490,C5493,C5494,C5495,C5496,C5497,C5499,C5500,C5501, C5502,C5503,C5507,C5508,C5509,C5510,C5511,C5512,C5515,C5516,C5517,C5518,C5519,C5520,C5521, C5522,C5524,C5525,C5526,C5527,C5528,C5529,C5530,C5532,C5533,C5534,C5535,C5536,C5537,C5538, C5539,C5544,C5545,C5546,C5547,C5549,C5550,C5553,C5555. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that results in the lens coefficients for an IOL model being downloaded from the Alcon server in an incorrect order.

What to do

Voluntary: Firm initiated · Telephone

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Alcon Research, Ltd.
Distributor
Worldwide distribution. US Nationwide and Argentina, Australia, Belgium, Brazil, Canada, China, Colombia, France, India, Japan, Mexico, Netherlands, Panama, Portugal, Romania, Spain, Thailand, UAE, and United Kingdom
Lot
Serial numbers: C5013,C5014,C5015,C5016,C5017,C5018,C5019,C5020,C5021,C5023,C5024,C5025,C5026,C5027,C5028,C5029,C5030,C5031,C5032,C5033,C5034,C5035,C5036,C5038,C5039,C5040,C5041,C5042,C5043,C5044, C5045,C5046,C5047,C5048,C5049,C5050,C5051,C5053,C5054,C5055,C5059,C5060,C5061,C5062,C5064, C5065,C5066,C5069,C5071,C5075,C5076,C5077,C5078,C5079,C5080,C5081,C5082,C5083,C5090,C5093, C5094,C5097,C5098,C5101,C5102,C5105,C5106,C5107,C5108,C5109,C5110,C5111,C5112,C5113,C5116, C5117,C5119,C5120,C5121,C5122,C5125,C5126,C5127,C5128,C5129,C5130,C5131,C5132,C5133,C5134, C5136,C5137,C5138,C5139,C5140,C5141,C5142,C5143,C5144,C5146,C5147,C5148,C5149,C5150,C5151, C5152,C5153,C5154,C5155,C5156,C5157,C5158,C5159,C5160,C5161,C5163,C5164,C5165,C5166,C5167, C5168,C5169,C5170,C5171,C5173,C5174,C5175,C5176,C5177,C5178,C5179,C5180,C5181,C5182,C5183, C5184,C5185,C5186,C5187,C5188,C5189,C5190,C5191,C5192,C5193,C5194,C5195,C5196,C5197,C5198, C5199,C5200,C5201,C5202,C5203,C5204,C5205,C5207,C5214,C5215,C5216,C5217,C5218,C5219,C5220, C5221,C5222,C5223,C5224,C5225,C5226,C5227,C5228,C5229,C5230,C5231,C5232,C5233,C5238,C5239, C5240,C5241,C5242,C5243,C5244,C5245,C5246,C5248,C5249,C5250,C5251,C5252,C5253,C5255,C5256, C5257,C5258,C5259,C5260,C5261,C5262,C5263,C5264,C5265,C5266,C5267,C5268,C5269,C5270,C5271, C5272,C5273,C5274,C5276,C5277,C5278,C5279,C5280,C5281,C5282,C5283,C5284,C5285,C5286,C5287, C5288,C5289,C5290,C5291,C5292,C5293,C5294,C5295,C5296,C5297,C5298,C5299,C5300,C5301,C5302, C5303,C5304,C5305,C5306,C5307,C5308,C5309,C5310,C5311,C5312,C5313,C5314,C5315,C5316,C5317, C5318,C5319,C5320,C5321,C5322,C5323,C5326,C5328,C5329,C5330,C5334,C5335,C5336,C5339,C5340, C5341,C5343,C5348,C5349,C5350,C5352,C5353,C5354,C5355,C5356,C5357,C5358,C5359,C5360,C5361, C5362,C5363,C5364,C5365,C5366,C5368,C5369,C5370,C5371,C5372,C5373,C5374,C5375,C5376,C5377, C5378,C5379,C5380,C5381,C5382,C5383,C5384,C5385,C5386,C5387,C5388,C5389,C5390,C5391,C5392, C5393,C5394,C5395,C5396,C5397,C5398,C5399,C5400,C5401,C5402,C5403,C5404,C5405,C5406,C5407, C5408,C5409,C5410,C5411,C5412,C5413,C5414,C5415,C5416,C5417,C5418,C5434,C5435,C5436,C5437, C5438,C5439,C5440,C5441,C5442,C5443,C5444,C5446,C5448,C5449,C5451,C5452,C5453,C5454,C5456, C5458,C5462,C5463,C5464,C5465,C5466,C5467,C5469,C5471,C5473,C5475,C5476,C5477,C5478,C5479, C5484,C5485,C5486,C5487,C5488,C5489,C5490,C5493,C5494,C5495,C5496,C5497,C5499,C5500,C5501, C5502,C5503,C5507,C5508,C5509,C5510,C5511,C5512,C5515,C5516,C5517,C5518,C5519,C5520,C5521, C5522,C5524,C5525,C5526,C5527,C5528,C5529,C5530,C5532,C5533,C5534,C5535,C5536,C5537,C5538, C5539,C5544,C5545,C5546,C5547,C5549,C5550,C5553,C5555
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ORA System with VerifEye+ Cart, Catalog Number 8065998307 For use during intraocular lens surgery recalled?

Yes. FDA published recall Z-3049-2017 on August 30, 2017. Reported hazard: Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that results in the lens coefficients for an IOL model being downloaded from the Alcon server in an incorrect order.

What should I do if I have this product?

Voluntary: Firm initiated · Telephone