CS 100 Intra-Aortic Balloon Pump. It supports the heart's left ventricle by increasing coronary perfusion and reducing left ventricular work. Coronary perfusion is increased by augmenting blood pressure during the diastolic phase of the cardiac cycle. This increase in aortic pressure promotes more blood flow through the coronary arteries. Left ventricular work is reduced by decreasing aortic end diastolic pressure and reducing resistance to ventricular ejection, resulting in a decrease in blood pressure during the systolic phase of the cardiac cycle.

Published September 13, 2017 · Recall initiated: July 17, 2017

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

False blood back detection alarm and the ingress of fluids into the IABP affecting various electronic circuit boards could potentially prevent initiation or continuation of therapy.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Maquet Datascope Corp - Cardiac Assist Division
Distributor
Worldwide Distribution - US Nationwide International: AFGHANISTAN, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BANGLADESH, BELARUS, BELGIUM, BOLIVIA, BOSNIA-HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, GUATEMALA, GUYANA, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOREA, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MALDIVES, MALTA, MAURITIUS, MEXICO, MOROCCO, MYANMAR, NEPAL, NETHERLANDS, NEW ZEALAND, NORTH KOREA, NORWAY, OMAN, PAKISTAN, PALESTINE, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TANZANIA, THAILAND, TRINIDAD, TUNISIA, TURKEY, TURKMENISTAN, UGANDA, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, UZBEKISTAN, VENEZUELA, VIETNAM & YEMEN.
Lot
0998-XX-3013-XX; 0998-UC-3013-XX
Risk classification
Class I