CS 100 Intra-Aortic Balloon Pump
Published August 2, 2017 · Recall initiated: June 16, 2017
Summary
FDA records medical device Class I recall Z-2737-2017, published August 2, 2017 and initiated June 16, 2017. Recalling firm: Maquet Datascope Corp - Cardiac Assist Division; Status: Terminated; Reported quantity: 12,319 units total; Lot: 0998-XX-3013-XX, 0998-UC-3013-XX. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device to initiate therapy.
What to do
Voluntary: Firm initiated · Press Release
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Maquet Datascope Corp - Cardiac Assist Division
- Distributor
- Worldwide Distribution - (in over 100 countries) US (Nationwide)
- Lot
- 0998-XX-3013-XX, 0998-UC-3013-XX
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACardiosave Rescue IABP, Part Number 0998-UC-0800-83November 7, 2018
- FDACS 100 Intra-Aortic Balloon Pump. It supports the heart's left ventricle by increasing coronary perfusion andSeptember 13, 2017
- FDACS 300 Intra-Aortic Balloon PumpAugust 2, 2017
- FDACS 100i Intra-Aortic Balloon PumpAugust 2, 2017
- FDADierbergs Kitchen, Santa Fe Grilled Chicken Wrap, NET WT. 11 oz, UPC 0 231789 804996August 2, 2017
- FDAMedtronic Visualase(R) Cooled Laser Applicator System (VCLAS) 15MM TIP, REF 9735561, (1) Laser Diffusing FiberAugust 2, 2017
Questions about this recall
Was CS 100 Intra-Aortic Balloon Pump recalled?
Yes. FDA published recall Z-2737-2017 on August 2, 2017. Reported hazard: The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device to initiate therapy.
What should I do if I have this product?
Voluntary: Firm initiated · Press Release