Coulter HmX CP and Coulter HmX AL Analyzers The purpose of the HmX Hematology Analyzer is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies. These studies include further measurements of cell size and cell distribution, biochemical investigation or any other test that helps diagnose the abnormality
Published September 13, 2017 · Recall initiated: June 26, 2017
Summary
FDA records medical device Class II recall Z-3114-2017, published September 13, 2017 and initiated June 26, 2017. Recalling firm: Beckman Coulter Inc.; Status: Terminated; Reported quantity: 3326 units; Lot: Catalog Number(s): HMX CP - 178832, 178833, 178834, A91062, A90994; HmX AutoLoader - 6605522, 6605523, 6605524; All software versions. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the LH 750, LH 780, LH 500 and HmX Analyzers may not flag or detect blasts in some blood samples.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Beckman Coulter Inc.
- Distributor
- Worldwide Distribution - US (nationwide) Internationally to Albania, Algeria, Andorra, Angola, Antigua and Barbuda, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Bulgaria, Burundi, Canada, China, Colombia, Congo, Croatia, Czech republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Germany, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, india, Indonesia, Italy, Japan, Iraq, Jordan. Kenya, Kuwait, Libya, Lithuania, Macau, Lebanon, Macedonia, Malawi, Malaysia, Mexico, Mongolia, Morocco, Mozambique, Netherlands, New Zealand, Nigeria, Pakistan, Panama, Paraguay, Philippines, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Swaziland, Sweden, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, UAE, UK, Uruguay, Venezuela, Vietnam, Virgin Islands, Yemen, Zambia, Zimbabwe.
- Lot
- Catalog Number(s): HMX CP - 178832, 178833, 178834, A91062, A90994; HmX AutoLoader - 6605522, 6605523, 6605524; All software versions.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520October 14, 2026
- FDADxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AUOctober 14, 2026
- FDABeckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU ModuleAugust 26, 2026
- FDABeckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE,August 26, 2026
- FDAAMLODIPINE BESYLATE TABLET, USP, 10 mg, 1000 count bottle, Rx Only, Manufactured by: Alkem Laboratories Ltd.,September 20, 2017
- FDAQualigen(R) The FastPack(R) IP Vitamin D Complete Immunoassay Kit, catalog number 25000068 Vitamin DSeptember 13, 2017
Questions about this recall
Was Coulter HmX CP and Coulter HmX AL Analyzers The purpose of the HmX Hematology Analyzer is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies. These studies include further measurements of cell size and cell distribution, biochemical investigation or any other test that helps diagnose the abnormality recalled?
Yes. FDA published recall Z-3114-2017 on September 13, 2017. Reported hazard: Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the LH 750, LH 780, LH 500 and HmX Analyzers may not flag or detect blasts in some blood samples.
What should I do if I have this product?
Voluntary: Firm initiated · Letter