Back to search
FDAFDA · Food and drugs·Recall number Z-3125-2024

Brand Name: PROBEAT Product Name: PROBEAT-FR Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-FR is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. Component: N/A

Published September 18, 2024 · Recall initiated: May 27, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are insufficient in number.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Hitachi America, Ltd., Particle Therapy Division
Distributor
U.S. Nationwide distribution in the states of TN, TX, and WA.
Lot
Lot Code: UDI:(01)04560333350204(11)240112(21)PTFR-01
Risk classification
Class II
Area (FDA)
device
Country
United States